OPERATING MANUAL FOR SELECT EDI 60
L300180 Iss:A
33
TECHNICAL SPECIFICATION
10.3
Product outputs
Make-up
rate (l/hr)
@ 10°C &
60psi
6
@ 25°C &
60psi
10
10.4
Treated water specification
•
Inorganics
> 1 M
Ω
.cm
•
pH
Neutral
•
Bacteria
> 99% rejection
•
Organics (TOC)
<50ppb
10.5
Water storage
Pure water storage capacity: 20 litres
10.6
Electrical specifications / connections
10.6.1
Mains supply
Electrical
Supply
KW
Rating
Current
Draw (Amps)
External Fuse
Rating
Single phase
100-230v
+10%,50/60
Hz and Earth
0.25
1.0
5 Amps
(conforming to
BS1362)
10.6.2
Fuse rating / type
IEC module fuse type: - (20x5mm), T5AH250V, conforming to IEC 127
Number per unit = 2
The mains supply is double pole/live & neutral fused.
10.6.3
Alarm port connection details
Application:
BMS Volt free alarm output.
Connector type:
Din 3-pin Plug (max rating 34Vdc / 24ac, 1 amp).
Maximum lead length= 10 metres.
Contact SUEZ for supply of alarm lead.