Function of Controls:
MODE/RESET Key: Press ONCE to reset Max/Min temperature record. When
the unit is sounding an alarm, press ONCE to turn off the beeping sound.
HIGH/LOW Key: Press ONCE to display high alarm setting. Press TWICE to
display low alarm setting. Press THREE TIMES to display current
temperature reading.
ALERT ON/OFF Key: When the unit is sounding an alarm, press ONCE to
turn off. However, the High-Alarm or Low-Alarm will still flash until the
temperature goes back to normal or the Alarm function is turned off.
Programming the Alarms:
The thermometer displays the internal temperature of the unit being
monitored, as well as the maximum and minimum temperatures reached. The
maximum and minimum readings reflect the highest and lowest temperatures
since the battery was inserted or since the MODE/RESET key was last
pressed. If the temperature rises above or falls below the set temperature
range, the alarm will sound. By default, the high alarm temperature setting is
8°C (46.4°F) and the low alarm temperature setting is 2°C (35.6°F). To reset
these limits, follow these instructions:
High Alarm Temperature Setting: Press the HIGH/LOW key ONCE, then press
MODE/RESET to enter the
High Alarm
temperature setting mode. Use the
Up and Down arrows to adjust the temperature accordingly. Press
MODE/RESET once again to confirm the setting.
Low Alarm Temperature Setting: Press the HIGH/LOW key TWICE, then
press MODE/RESET to enter the
Low Alarm
temperature setting mode.
Use the Up and Down arrows to adjust the temperature accordingly. Press
MODE/RESET ONCE to confirm the setting.
By default, the low and high alarms should be audible. To turn alarms on/off:
Audible High Alarm Setting: To make sure the high alarm is audible, press
HIGH/LOW key ONCE (to activate HI-SET on the screen). Then press the
ALERT ON/OFF key ONCE to activate the HIGH-ALARM beeping sound.
Press again to deactivate.
Audible Low Alarm Setting: To make sure that the low temperature alarm is
audible, press the HIGH/LOW key TWICE to activate LOW-SET on the
screen. Then press the ALERT ON/OFF key ONCE to activate the LOW-
ALARM beeping sound. Press again to deactivate.
NOTE: For an activated high alarm and low
alarm buzzer, the display should indicate
HI-ALARM and LO-ALARM as shown here.
More details about the features and usage of the Thermometer/ Alarm can be
found in the User Manual accompanying the unit.
THERMOMETER / ALARM
(cont’d)
Felix Storch, Inc.
Page 11