background image

18

19

Guidance and manufacturer’s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environments listed below, and should only be 

used in such environments:

Immunity 

test

IEC 60601 test 

level

Compliance level

Electromagnetic environment 

-guidance

Conducted RF

IEC 61000-4-6

3V rms

At 0.15-80MHz

6V rms

At ISM & Radio

Amateur Freq

3V rms

At 0.15-80MHz

6V rms

At ISM & Radio

Amateur Freq.

Portable and mobile RF communications 

equipment should be used no closer to 

any part of the device, including cables, 

than the recommended separation 

distance calculated from the equation 

applicable to the frequency of the 

transmitter.

Radiated RF

IEC 61000-4-3

Proximity

elds from RF

wireless 

communications

equipment

IEC 61000-4-3

10 V/m at 

80 - 2700MHz AM 

Modulation And 

9 - 28V/m at

385 - 6000MHz, 

Pulse Mode and 

other Modulation. 

The system shall be 

tested as specied in

IEC60601-1-2 

Table 9

for proximity elds 

from RF wireless 

communications

equipment using 

the test methods 

specied in IEC 

61000-4-3

10 V/m at 

80 - 2700 MHz AM 

Modulation And 

9 - 28V/m at

385 - 6000MHz, 

Pulse Mode and 

other Modulation. 

The system shall be 

tested as specied in

IEC60601-1-2 

Table 9

for proximity elds 

from RF wireless 

communications

equipment using 

the test methods 

specied in IEC 

61000-4-3

Recommended separation distance

Considering to reduce the minimum 

separation distance, based on RISK 

MANAGEMENT, and using higher 

IMMUNITY TEST LEVELS that are 

appropriate for the reduced minimum 

separation distance. Minimum  

separation distances for higher 

IMMUNITY TEST LEVELS shall be 

calculated using the following equation:

E=6/d

where P is the maximum power in W, 

d is the minimum separation distance 

in m, and E is the IMMUNITY TEST 

LEVELS in V/m. Field strengths from xed 

RF transmitters, as determined by an 

electromagnetic site survey, 

a

 should 

be less than the compliance level in 

each frequency range.

b

 Interference 

may occur in the vicinity of equipment 

marked with the 

following symbol:

NOTE 1

  At 80MHz and 800MHz, the higher frequency range applies.

NOTE 2

  These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

a

 Field strengths from xed transmitters, such as base stations for radio (cellular / cordless) telephones 

and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be 

predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF 

transmitters, an electromagnetic site survey should be considered. If the measured eld strength in 

the location in which the device is used exceeds the applicable RF compliance level above, the device 

should be observed to verify normal operation. If abnormal performance is observed, additional 

measures may be necessary, such as reorienting or relocating the device.

b

 Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.

P

Test specications for enclosure port immunity to RF wireless communications equipment.

Test frequency (MHz)

Modulation

IMMUNITY TEST LEVEL (V/m)

385

Pulse modulation 18Hz 

a)

27

450

FM ±5kHz deviation 1kHz sine 

b)

28

710

745

780

Pulse modulation 217Hz 

a)

9

810

870

930

Pulse modulation 18Hz 

a)

28

1720

1845

1970

Pulse modulation 217Hz

 a)

28

2450

Pulse modulation 217Hz

 a)

28

5240

5500

5785

Pulse modulation 217Hz 

a)

9

NOTE:

If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and 

the ME EQUIPMENT or ME SYSTEM may be reduced to 1m, The 1m test distance is permitted by IEC 

61000-4-3.

a)

 The carrier shall be modulated using a 50% duty cycle square wave signal.

b)

 AS an alternative to FM modulation, 50% pulse modulation at 18Hz may be used because while it   

does not represent actual modulation, it would be worst case.

Summary of Contents for 16994

Page 1: ...1 Upper Arm Blood Pressure Monitor with Voice Broadcast Technology Instruction Manual Model HL858A2 LT Item 16994 Manual Version 001 Issue Date XX ...

Page 2: ...curate idea of your blood pressure MEASUREMENT METHOD HL858A2 LTAutomatic Upper Arm Blood Pressure Monitor measures blood pressure and heart rate by oscillometric method meaning the fluctuations in pressure are measured Once the cuff is wrapped around your upper arm just turn on the monitor and inflation automatically starts The inflation of the cuff creates pressure around the arteries inside upp...

Page 3: ...osis diabetes liver disease kidney disease severe hypertension peripheral circulation please consult your healthcare professional before using the device Blood pressure measurements taken with this device are equivalent to those obtained by a trained observer using the cuff stethoscope auscultation method and are within the accuracy limits prescribed by the Standard of EN 1060 4 If the cuff is wor...

Page 4: ...perly BP Category Indicator The arrowhead points out the specific BP Category that your measurement reading fits in Average of Last 3 Measurements ThissymbolappearswhenLCDdisplaysaverage value of last3readings IRREGULAR HEARTBEAT DETECTOR The symbol will appear on screen indicating a certain heartbeat irregularity was detected during measurement The heartbeat rhythm that is more than or less than ...

Page 5: ... Insert 4 new AAA 1 5 V LR03 alkaline batteries into the compartment as shown Make sure the polazrities and align with the symbols inside the compartment 3 Replace the batteries when The low battery icon appears The display is dim or does not turn on Do not mix new and used batteries Do not mix alkaline standard carbon zinc or rechargeable cadmium batteries Such action may shorten the battery life...

Page 6: ...uff inanyway MEMORY FUNCTION 1 Themonitorfeaturesmemorystoragewithdateandtimestampsfortwo 2 users saving 120recordsperuser 240total 2 Aftereachmeasurement thesystolicanddiastolicpressure heartrateandBPCategory Indicator IrregularHeartbeatDetectorandExcessivebodymotiondetector if any withthe dateandtimewillbeautomaticallystored Eachnewreadingwillreplacetheoldest TAKING THE MEASUREMENT 1 With the cu...

Page 7: ...ol The cuff was wrapped incorrectly for example too loosely or too tightly Please reference applying the Cuff section to wrap the cuff correctly Note If EP appears on the display just return the device to your local distributor or importer RECALLING DATA 1 Press buttontoselectUser1or2 2 PressM buttontoenterMemoryMode If thereisnodatastored nothing exceptmonth date andtime willappearonthedisplay If...

Page 8: ...hthelimitsforaClassBdigital device pursuanttoPart15of theFCCRules Theselimitsaredesignedtoprovidereasonable protectionagainstharmfulinterferenceinaresidentialinstallation Thisequipmentgenerates usesandcanradiateradiofrequencyenergyand if notinstalledandusedinaccordancewith theinstructions maycauseharmfulinterferencetoradiocommunications However there isnoguaranteethatinterferencewillnotoccurinapar...

Page 9: ...15kV air discharge In the case of air discharge testing the climatic conditions shall be within the following ranges AmbientTemperature 15 C 35 C Relative Humidity 30 60 Power frequency 50 or 60Hz magnetic eld IEC 61000 4 8 30A m 50 or 60Hz 30A m 50 or 60Hz Powerfrequency magnetic eldsshouldbeatlevels characteristicof atypical locationinatypicalcommercial orhospitalenvironment Electrical fast tran...

Page 10: ...ed by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures object...

Page 11: ...399 8 30 am 5 00 pm EST Monday Friday or contact us at info stareliteinc com Our customer service will be happy to assist you 2021 Sunbeam Products Inc All rights reserved Distributed by Star Elite Inc Montreal Canada H3B 3X9 SE005 080621 Upper Arm Blood Pressure Monitor Model HL858A2 LT Printed in China ...

Reviews: