Position the artery mark over the main artery (on the inside of your 

arm). 

Tip:

 To locate the main (brachial) artery, press 2 fingers 

approximately 1” (2-3cm) above the bend of your elbow on the 

inside of your left arm. Identify where the pulse can be felt the 

strongest.  Fit the cuff snugly, leaving enough space for 1” (2 - 3 cm) 

between the inner elbow and the lower edge of the cuff. If the cuff is 

positioned correctly, the hook-and-loop fabric will be on the outside 

of the cuff and the metal ring will not touch your skin. The bottom 

edge of the cuff should be located within the Proper Fit Range. 

5.  Pull the end of the cuff so it tightens evenly around your arm.  

6.  Make sure the cuff is snug enough to stay in place during 

measurement, but not too tight. You should be able to insert one 

finger between the cuff and your arm. 

7.  Press the hook-and-loop fabric strip to secure. 

8.  Plug the arm cuff tube into the monitor port.

TIPS FOR ACCURATE MEASUREMENT

 

1.  Avoid taking blood pressure: 

•  within one (1) hour of eating or drinking (especially caffeinated beverages) 
•  immediately after smoking
•  within 20 minutes of bathing
•  while talking or moving your hands/fingers 
•  in a cold environment 
•  when you feel the urge to urinate 

2.  Take the measurement in a silent room. 
3.  Sit comfortably in an upright chair with back supported, feet flat on the floor and legs 

uncrossed. 

4.  Rest your testing arm on a flat surface with the palm turned upward and the cuff level 

with the right atrium of the heart. 

5.  Rest and relax for 5 minutes before measuring. Take 5 - 6 deep breaths.  
6.  Stay quiet and still during measuring. 
7.  Wait at least three minutes between measurements. 
8.  For meaningful comparison, try to measure under similar conditions. For example, take 

daily measurements at approximately the same time, on the same arm, or as directed 
by a physician. 

IMPORTANT

 

•  Make sure there are no kinks in the air tube. 
•  Do not twist the cuff in any way. 

MEMORY FUNCTION 

 

1.  The monitor features memory storage with date and time stamps for two (2) users, saving 

120 records per user (240 total).  

2.  After each measurement, the systolic and diastolic pressure, heart rate and BP Category 

Indicator, Irregular Heartbeat Detector and Excessive body motion detector (if any) with the 
date and time will be automatically stored.  Each new reading will replace the oldest. 

TAKING THE MEASUREMENT 

 

1.  With the cuff wrapped around your arm, press 

START / STOP

 

button to begin. 
•  All display symbols will appear on the screen for 1.5 seconds, 

followed by “00” to indicate “Ready to Measure.”  

2.  The cuff will automatically inflate. 

•  During inflation, the monitor detects your blood pressure and 

pulse rate.

•  The Heartbeat Symbol         flashes at every heartbeat.

3.  Do not move or talk until the entire measurement process is 

completed. 

4.  The cuff automatically deflates when your blood pressure and heart rate measurement 

is complete. 
•  Your systolic rate, diastolic rate, heart rate, corresponding BP Category Indicator, 

Irregular Heartbeat Detector and Excessive body motion detector (if any) are 
displayed with date and time for one (1) minute and results saved to memory 
automatically. 

•  If Voice Broadcast is ON (default setting), the device will announce your test results 

out loud. 

•  The device automatically shuts off after one minute of no use.

NOTE: 

If the cuff does not stop inflating, remove the cuff at once. To stop measurement at 

any time, press 

START / STOP

. The cuff will deflate immediately after the button is pressed.

•  If the cuff inflates without stopping, unwrap the cuff immediately. 
•  Do not press the 

START / STOP

 button if the cuff is not wrapped around your upper arm. 

•  Do not wrap the cuff around any body part other than your arm. 
•  Do not use this device on a wounded or injured arm. Do not use this device after recent 

surgery on the wrist or arm (doing so may increase the risk of infection at the surgical site). 
Wait until after the wound has healed to use this device. 

•  If you have an infectious skin condition or disease or the device is used by others with 

infectious skin condition or disease, dispose of the device without reuse. 

•  Before use, always inspect the device. If you notice blood or other soil on cuff, do not reuse.

Summary of Contents for 16994

Page 1: ...1 Upper Arm Blood Pressure Monitor with Voice Broadcast Technology Instruction Manual Model HL858A2 LT Item 16994 Manual Version 001 Issue Date XX ...

Page 2: ...curate idea of your blood pressure MEASUREMENT METHOD HL858A2 LTAutomatic Upper Arm Blood Pressure Monitor measures blood pressure and heart rate by oscillometric method meaning the fluctuations in pressure are measured Once the cuff is wrapped around your upper arm just turn on the monitor and inflation automatically starts The inflation of the cuff creates pressure around the arteries inside upp...

Page 3: ...osis diabetes liver disease kidney disease severe hypertension peripheral circulation please consult your healthcare professional before using the device Blood pressure measurements taken with this device are equivalent to those obtained by a trained observer using the cuff stethoscope auscultation method and are within the accuracy limits prescribed by the Standard of EN 1060 4 If the cuff is wor...

Page 4: ...perly BP Category Indicator The arrowhead points out the specific BP Category that your measurement reading fits in Average of Last 3 Measurements ThissymbolappearswhenLCDdisplaysaverage value of last3readings IRREGULAR HEARTBEAT DETECTOR The symbol will appear on screen indicating a certain heartbeat irregularity was detected during measurement The heartbeat rhythm that is more than or less than ...

Page 5: ... Insert 4 new AAA 1 5 V LR03 alkaline batteries into the compartment as shown Make sure the polazrities and align with the symbols inside the compartment 3 Replace the batteries when The low battery icon appears The display is dim or does not turn on Do not mix new and used batteries Do not mix alkaline standard carbon zinc or rechargeable cadmium batteries Such action may shorten the battery life...

Page 6: ...uff inanyway MEMORY FUNCTION 1 Themonitorfeaturesmemorystoragewithdateandtimestampsfortwo 2 users saving 120recordsperuser 240total 2 Aftereachmeasurement thesystolicanddiastolicpressure heartrateandBPCategory Indicator IrregularHeartbeatDetectorandExcessivebodymotiondetector if any withthe dateandtimewillbeautomaticallystored Eachnewreadingwillreplacetheoldest TAKING THE MEASUREMENT 1 With the cu...

Page 7: ...ol The cuff was wrapped incorrectly for example too loosely or too tightly Please reference applying the Cuff section to wrap the cuff correctly Note If EP appears on the display just return the device to your local distributor or importer RECALLING DATA 1 Press buttontoselectUser1or2 2 PressM buttontoenterMemoryMode If thereisnodatastored nothing exceptmonth date andtime willappearonthedisplay If...

Page 8: ...hthelimitsforaClassBdigital device pursuanttoPart15of theFCCRules Theselimitsaredesignedtoprovidereasonable protectionagainstharmfulinterferenceinaresidentialinstallation Thisequipmentgenerates usesandcanradiateradiofrequencyenergyand if notinstalledandusedinaccordancewith theinstructions maycauseharmfulinterferencetoradiocommunications However there isnoguaranteethatinterferencewillnotoccurinapar...

Page 9: ...15kV air discharge In the case of air discharge testing the climatic conditions shall be within the following ranges AmbientTemperature 15 C 35 C Relative Humidity 30 60 Power frequency 50 or 60Hz magnetic eld IEC 61000 4 8 30A m 50 or 60Hz 30A m 50 or 60Hz Powerfrequency magnetic eldsshouldbeatlevels characteristicof atypical locationinatypicalcommercial orhospitalenvironment Electrical fast tran...

Page 10: ...ed by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures object...

Page 11: ...399 8 30 am 5 00 pm EST Monday Friday or contact us at info stareliteinc com Our customer service will be happy to assist you 2021 Sunbeam Products Inc All rights reserved Distributed by Star Elite Inc Montreal Canada H3B 3X9 SE005 080621 Upper Arm Blood Pressure Monitor Model HL858A2 LT Printed in China ...

Reviews: