• The stand up procedure automati
c
ally locks the front and rear cast
e
r
arm suspension.
WARNING!
• Before using the standing function ensu
r
e that the front knee fi
x
ation
is in place.
WARNING!
• The chest strap may only be assembled and adapted by an authorized
dealer.
CAUTION!
•
To prevent choking hazards the belt may not be adjusted too high
or too low relative to the users body.
to pressure from the chest strap.
•
Never use the chest strap as a patient restraint
.
•
Never use the chest strap on a user who is comatose or agitated.
DANGER!
Fig.
7
.
16
247556 Rev. A
29
Q700-UP M
7
.
8
Stand Up functio
n
• If the locking procedure fails the chair will automati
c
ally stop in a
safe
position. The se
a
t can be returned to the normally seatedposition in
a
n
y mode.
• Only use the standing function on a sound and l
e
vel surface.
Ensure the belt position does not i
n
terfere with any other device
a
tt
ached to the wheelchair.
This power chair has an electric standing function. This function m
ak
es
it possible for the user to stand up in the wheelchair.
7.8.1 Using a chest stra
p
• Before operating the
s
tand-up function the che
s
t strap and the
positioning belt, supplied as
s
tandard has to be buckled in
accordance
with these user instructions.
• Always also use the knee fi
x
ation
f
or the stand-up procedure.
DANGER!
Due to the risk of choking it is extremely hazardous to use the
belts without using the knee fi
x
ation.
Chest strap Adjustments:
Setting the height of the chest strap (Fig. 7.16).
NOTE:
A 5mm Hexagon wrench, (Allen Key), is required.
Positioning
• Seat the user into the chair in the correct seating position.
• Loosen the Allen screw to be able to position the chest strap at th
e
•
Make sure the rider does not slide in the wheelchair seat. If this
occurs, the rider may suffer chest compression or suffocate due
desired height.
• Tighten the Allen screws to a torque value of 5 to 5.8 ft-lbs
(6.8 to 7.9 Nm).