Everyday Activity Seat User Manual
Before re-issuing a product, we
recommend it is disinfected using the
following procedure:
Note: Always clean the product first to
remove accumulated soiling prior to
disinfecting/sanitising
SECTION 16:
RE-ISSUING LECKEY
PRODUCTS
Most Leckey products are assessed
and ordered to meet the needs of
an individual user. Before reissuing
a product we recommend that the
therapist prescribing the product has
carried out an equipment compatibility
check for the new user and has
ensured that the product being
reissued contains nomodifications or
special attachments.
A detailed technical inspection should
be carried out on the product prior to
reissuing. This should be carried out
by a technically competent person
who has been trained in the use and
inspection of the product. Please refer
to
section 15 for the required checks to
be carried out.
1. Where possible, we recommend
the use of an automated “fogger”
disinfection system that is used
to completely cover the product
surfaces with a disinfectant agent.
The manufacturer’s instructions
for the chosen system should
be referred to prior to use and
followed exactly.
2. Where an automated disinfection
system is not available, all product
surfaces should be wiped over
with disinfectant surface wipes or
liquid disinfectant on a soft cloth.
Suitable products are 70% IPA,
70% Alcohol or diluted bleach. The
manufacturer’s instructions for the
chosen product should be followed.
Product upholstery and fabrics
should be cleaned thoroughly in
accordance with section, Cleaning
and Care Information, of this
manual.