MELODY SERVICE MANUAL
PAGE 7
2006
SUNRISE MEDICAL
Section 3
Check that the batteries are fully charged and in good
condition; and check all cables and connections. Check the
connections to the left motor, look for a loose or damaged
connector.
Measure the resistance from the bottom contact of the red
thick wire on the 4-pin left motor connector to each of the
top contacts of the connector see (figure A3.3.1). Measure
the resistance from the bottom contact of the black thick
wire on the 4-pin left motor connector to each of the top
contacts of the connector see (below right). If all of the
readings are open, then replace the controller. If any of the
readings are short, then replace the left motor.
3.3 Three Bars - Left Motor Wiring Trip
A3.3.1
Test 1
Test 2
Test 3
Test 4
Understanding Controller Diagnostics Codes (cont.)
If any of the readings are short, then replace the left motor.
If all of the readings are open, then replace the controller.