MELODY SERVICE MANUAL
PAGE 8
2006
SUNRISE MEDICAL
Section 3
Understanding Controller Diagnostics Codes (cont.)
Check that the batteries are fully charged and in good
condition; and check all cables and connections. Check the
connections to the right motor, look for a loose or damaged
connector.
Use the meter to check the resistance across the two
bottom contacts of the thicker wires on the 4-pin motor
connector as shown in (figure A3.4.2). If the meter reads
between 0 to 1.5 ohms, then replace the controller. If none
of the above corrects the problem, replace the right motor.
Otherwise, check the brushes on the right motor. Ensure
that they are not excessively worn, (replace as required) as
shown in (Figure A3.4.1).
3.4 Four Bars- Right Motor Disconnected
A3.4.1
A3.4.2
If the meter reads between 0 and 1.5 ohms, then replace the controller. If this does not correct the problem,
then replace the right motor.