MELODY SERVICE MANUAL
PAGE 3
2006
SUNRISE MEDICAL
Troubleshooting: No Power (cont.)
1.5 Battery Connection Test
Check that the female VR2 Bus plug on the chair has volt-
age. Set the meter to dc volts and measure pins 4 (using
the red lead of the meter) and 1 (using the black lead of the
meter) as shown in (figure A1.5.1)
A1.5.1
A1.5.2
If the voltage meter reads full voltage, then replace the controller.
If the voltage meter reads full voltage, then replace the joy-
stick module
If the voltage meter reads zero voltage measure the cor-
responding pins on the VR2 controller as shown in (figure
1.5.2).
If the voltage meter reads full voltage, then replace the
jumper cable. If the voltage meter reads zero, then mea-
sure the Battery Connector as shown in (fig 1.5.3)
If the voltage meter reads full voltage, then replace the con-
troller, or else proceed to the next step.
A1.5.3