Model CR1200R Digital Signal Analyzer
Page 44
menu Screen 2, with the functions: Knob Ref / Knob Mkr1 / Knob
Mkr2 / Knob Span / More (Figure 3-27)
K
NOB
R
EFERENCE
Selecting the KNOB REF (F1) key highlights the current reference level.
Once the KNOB REF button has been selected, the knob can then be used to
change the reference level. The reference level can be set at any level from
+50 dBmV to –10 dBmV in 4 dB increments.
KNOB MKR1 & KNOB MKR2—M
ARKERS
The CR1200R has two markers available. Each marker can be
scrolled across the full display showing both timing and level infor-
mation as well as the time and amplitude delta between the two mark-
ers. When KNOB MKR1 (F2) is pressed, the knob controls Marker
1. When KNOB MKR2 (F3) is pressed, the knob controls Marker 2.
The display area on the left side of the screen provides the measure-
ment results for the markers.
K
NOB
S
PAN
The time span is displayed at the top of the screen. The span may be
set by pushing the KNOB SPAN (F4) key. When this button has been
pressed, turning the knob will change the span setting. The selections
are 100 µSec to 500 mSec. The tuning knob continues to control the
SPAN until another function is selected.
M
ORE
(S
PECTRUM
A
NALYZER
S
CREEN
TWO
)
Pressing MORE (F5) from the second ZERO SPAN screen (Figure 3-
31) takes the user back to the initial ZERO SPAN screen (Figure 3-
26).