3.1 Environmental Operating Conditions
• Ambient Temperature: 41°F – 86°F (5°C – 30°C)
• Relative humidity: 5% – 95%
• Altitude: less than 6,561 feet (2,000 meters)
3. Additional Requirements
2.4 Errors and Troubleshooting
Consult your CPAP machine user manual.
Rinsing Instructions
Following removal from CIDEX® OPA Solution, thoroughly rinse the
medical device by immersing it completely in a large volume (e.g. 8
liters) of sterile water.
• Keep the device totally immersed for a minimum of 1 minute.
• Manually flush all lumens with large volumes (not less than
100ml) of rinse water.
• Remove the device and discard the rinse water. Always use
fresh volumes of water for each rinse. Do not reuse the water
for rinsing or any other purpose.
• Repeat the procedure TWO (2) additional times, for a total of
THREE (3) RINSES, with large volumes of fresh water to remove
CIDEX® OPA Solution residues.
• Refer to the disinfectant solution manufacturer’s instructions
for use for additional information and cautions.
Caution: Follow disinfectant solution manufacturer’s instructions for
use. Failure to do so may result in inadequate disinfection.
Review all manufacturer cautions and warnings.