80-0084-00-RevB SunTech Vet30E User Manual | 38
2019-12-04
AccuVet Temperature Probe, 2m
52-0022-00
Temperature probe, 6.5 ft (2 m)
AccuVet Temperature Probe, 3m
52-0023-00
Temperature probe, 9.8 ft (3 m)
Masimo M-LNCS
TM
YI AH SpO
2
w/ 3 clips
52-0024-00
Y-lingual sensor, 3 ft (0.9 m)
Masimo M-LNCS TF-I AH Transflectance SpO
2
Sensor
52-0026-00
Reflectance sensor, 3 ft (0.9 m)
Masimo M-LNCS Patient Cable
52-0012-01
Patient cable, 10 ft (3 m)
Masimo AH SpO
2
Replacement Clips (3)
52-0025-00
Pack of 3 sizes (small, medium, large)
Rechargeable Lithium-Ion Battery
17-0026-00
7.2 V, 2.2 Ah, 15.8 Wh
10.6 Status Messages
If the SunTech Vet30E has a problem performing a task, a short beep will occur (if speakers are ON), the
START/STOP button will turn Red and a Status Message will appear on the monitor screen. Take action as directed
on the screen, or as suggested in the table below.
Status Message
Reason
Solution
Air Leak
There is a leak in the cuff, hose
or monitor. Also possible if cuff
or hose is not attached to the
monitor.
Check that the hose is connected to the monitor and cuff.
Check that the cuff is properly tightened.
Check that the cuff is in the correct position.
Check that the cuff is not leaking air.
Check that the hose connections are not damaged or loose.
Value out of
Range
The blood pressure value is
outside of SunTech Vet30E
’s
published ranges.
Make sure you are using the proper cuff size.
Patient may have been moving too much.
Check that the cuff is properly tightened and in proper
position.
Poor Signal
Quality
Monitor is not receiving a strong
signal from patient. Also
possible if a rapid deflation
occurs during a measurement.
Check that cuff is in correct position.
Check the patient.
Check that cuff is properly tightened.
Check that correct size cuff is used.
Artifact Detected Monitor is picking up on
unexpected noise or
movements.
Check animal mode. It might be on wrong setting.
Check patient for motion, trembling. Too much movement.
Check that cuff is in correct position.
Check that correct size cuff is used.
Measurement
Too Long
The monitor is not detecting
strong and consistent signals
from the patient for an extended
time period.
Check to ensure cuff is fitting snugly on patient and is
positioned properly.
Check that cuff is in correct position.
Check patient for moving, trembling.
Check Batteries! Battery power is low.
See the battery life indicator on the Main Screen.
Recharge or replace battery as needed.
Air Blockage
Air is not able to pass through
the hose or cuff properly.
Check that hose has no sharp bends and is not pinched.
Check that patient is not lying or stepping on cuff.