80-0084-00-RevB SunTech Vet30E User Manual | 44
2019-12-04
Subsequent initial inflation: previous Sy 30 mmHg
Transducer Accuracy:
±3 mmHg between 0 mmHg and 300 mmHg
Transducer Calibration:
Recommended bi-annually or if a calibration problem is suspected
Temperature Specifications
Temperature Range:
26
⁰
C to 46
⁰
C
Temperature Accuracy: ±0.3
⁰
C plus the temperature sensor tolerance
Temperature Resolution: 0.1
⁰
C
Sensor:
YSI 400 compatible
Display Update Rate:
1 Hz
Measurement Mode:
Direct measurement
Transient Measurement Time: 45 seconds
Masimo Pulse Oximetry Specifications
Accuracy specifications are statistically distributed, and only about two-thirds of the measurements fall within the 1
Std. Dev. specification.
M-LNCS YI Multisite Sensor
M-LNCS TF-I Reflectance Sensor
SpO
2
Range
0-100%
0-100%
SpO
2
Accuracy
±2% at 70%-100%, for patients
weighing >30 kg
Undefined, for patients weighing
<30 kg
±2% at 70%-100%, for patients
weighing >30 kg
Undefined, for patients weighing
<30 kg
Pulse Rate Range
25-240 BPM
25-240 BPM
Pulse Rate Accuracy
±3 BPM, for patients weighing >30
kg
Undefined, for patients weighing
<30 kg
±3 BPM, for patients weighing >30
kg
Undefined, for patients weighing
<30 kg
SpO
2
Averaging Time:
8 second averaging
TF-I Sensor Wavelength: Red 660nm
Infrared 880nm
Ingress Protection:
IPX1
AccuVet Pulse Oximetry Specifications
SpO
2
Range:
0-99%
SpO
2
Accuracy:
±2% at 70%-99%
<70% unspecified
SpO
2
Resolution:
1% increments
SpO
2
Averaging:
8 pulse beat average
Calibration:
Factory calibrated over the range of 70% to 99% SpO2 using human blood samples to
functional saturation. Test methods available upon request. No in-service calibration is
required.
Sensor Wavelength:
Red 660nm, 2mW (typical)
Infrared 905nm, 2-2.4mW (typical)
Display Update Rate:
1 Hz (maximum age of SpO2 and pulse rate data is 35 seconds)
Pulse Oximetry Units:
% SpO
2
Pulse Rate Range:
18-400 BPM
Pulse Rate Resolution:
1 BPM
Pulse Rate Units:
Beats per minute
Pulse Rate Accuracy:
±2% or 2 BPM, whichever is greater
Pulse Rate Averaging:
8 second average
Ingress Protection:
IPX1
AccuVet Reflectance Sensor Specifications
SpO
2
Range:
50-100%