80-0084-00-RevB SunTech Vet30E User Manual | 8
2019-12-04
CAUTION:
Electrical shock and flammability hazard: Before cleaning, always turn off the device and disconnect from
any power source.
CAUTION:
Do not place the monitor on electrical equipment that may affect the device, preventing it from working
properly.
CAUTION:
To ensure that alarm limits are appropriate for the patient being monitored, check the limits each time the
monitor is used.
CAUTION:
Electrical Shock Hazard: Carry out periodic tests to verify that leakage currents of patient-applied circuits
and the system are within acceptable limits as specified by the applicable safety standards. The summation of
leakage currents must be checked and in compliance with IEC 60601-1 and UL60601-1. The system leakage current
must be checked when connecting external equipment to the system. When an event such as a component drop of
approximately 1 meter or greater or a spillage of blood or other liquids occurs, retest before further use. Injury to
personnel could occur.
CAUTION:
Disposal of product - Comply with local laws in the disposal of the device and/or its accessories.
CAUTION:
To minimize radio interference, other electrical equipment that emits radio frequency transmissions
should not be in close proximity to the pulse oximeter.
2.
Icons and Symbols
Some of the symbols, listed in the table below refer to the following FDA SDO Consensus standards:
•
Recognition #5-103, ISO 7000: 2014: Graphical symbols for use on equipment - Registered symbols
•
Recognition #5-116, ISO 7010: 2011: Graphical symbols - Safety colours and safety signs - Registered
safety signs
•
Recognition #5-102, ISO 60417: 2002 DB: Graphical symbols for use on equipment
•
Recognition #5-117, ISO 15223-1: 2016: Medical devices - Symbols to be used with medical device labels,
labelling, and information to be supplied - Part 1: General requirements
Symbol
Description
Standard/Source
General Warning Sign
ISO 7010-W001
Batch Code
ISO 7000-2492
Caution
ISO 7000-0434A
Classification
–
Class II
IEC 60417-5172
Refer to Instruction Manual
ISO 7010-M002
Authorized representative in the European Community
SunTech Design
USB-A or USB-B
Industry
Direct Current
IEC 60417-5031
Alternating Current
IEC 60417-5032
Polarity of DC power connector
IEC 60417-5926
Rechargeable battery
IEC 60417-5639
Consult Instructions for Use
ISO 7000-1641
This product meets the requirements of the applicable
Directives
EU Directive
Disposal in compliance with WEEE Directive
WEEE Directive