3 Installation
3.1 Initial Inspection
3.2 Installation
3.2.1 Mounting the Delayline Detector
Visual inspection of the system is required to ensure that no damage has occurred during shipping.
Should there be any signs of damage, please contact your provider immediately. Please check the delivery
according to the packing list (
) for completeness.
• Delayline Detector Unit (delivered in a vacuum container)
• Pulse Processing Unit ACU
• 1 x SHV Termination Plug
• 1 x DLD Readout Cable (HDMI) with Strain Relief
• CD with Documentation and Software
Table 1: Packing list for the DLD8080 R4.30 & R4.31 Detector.
The detector is transported under vacuum. Proceed as follows, to install the detector into your vacuum
chamber:
• Vent the transport container carefully and slowly (use N2 for venting if possible).
• Remove the transport container carefully.
• Before installing the detector to your chamber, check the front side of the MCP stack for particles.
The microchannel plates in front of the detector should be protected from exposure
to particle contamination. Particles that stick to the plate can be removed by carefully
using a single-hair brush carefully and/or with dry nitrogen. Reading the instructions
“microchannel plates” in
is strongly recommended.
• Install the detector into your vacuum chamber.