9
DLD8080 R4.30 & R4.31 Manual
• Keep the transport container in case that the detector must be sent back for repair. It can also be
used for storing the detector temporarily (not longer than one month) when not installed in a vacuum
chamber.
Note
The detector should be kept under vacuum all the time.
The best long term storage environment is an oil free vacuum in the 10E-2mbar range or
better.
The transport container is not designed for a long term storage of the detector. A storage
time of one month should not be exceeded.
The transport housing should be evacuated regularly with a period of at least one month,
in case that the transport housing must be used for a longer storage time.
3.2.2 Detector Orientation
marks the 0/0 position of the DLD image which corresponds to the upper left
corner of the image in the GUI software.
Figure 1: 0/0 position of the DLD image (black dot).
Image 0/0 position
3.2.3 Cabling and High Voltage
• The general connection scheme of the Delayline Detector including its readout package is shown in
.
DLD8080 R4.30 & R4.31 Manual | Surface Concept GmbH