background image

                                                                                                   MCPD18 R104 

    

It is not unusual that the operation voltage must be increased several times especially in the
first year of operation.

Typically,   the   final   operation   voltage   at   the   end   of   the   burn-in   period   will   be   close   to   the   recommended
maximum   operation   voltage.   An   increase   of   the   operation   voltage   above   the   recommended   maximum
operation voltage is in principal possible and allowed, as long as there is no significant increase in the dark
count rate, bright spots at the detection area and/or appearance of high voltage sparking.

The operation voltage can be increased  even above the recommended maximum operation
voltage. A safe and reliable operation will still be possible as long as there is no increase in
the dark count rate and/ or no appearance of high voltage sparking.

It   can   be   sufficient   to   estimate   the   correct   operation   voltage   by   monitoring   the   detector   results,   when
increasing   the   voltage   to   compensate   the   gain   degradation.   The   exact   operation   voltage   can   always   be
determined by measuring a so called MCP curve. The MCP curve plots the detected count rate as a function
of the operation voltage. Do as follows to measure an MCP curve:

Start the detector operation with a homogeneous illumination (as good as possible).

Use an appropriate device to measure the count rate output of the detector.

Decrease the operation voltage until the count rate of the detector is reduced to zero.

Increase  the operation   voltage   in  steps   of 50V   until  far  above  the  last  operation  voltage  (please
respect   the   recommended   maximum   operation   voltage)   or   until   the   detector   shows   significant
artifacts. Measure the count rate for each voltage step.

Plot the measured count rate as a function of the operation voltage. This is the MCP curve. It should
show a change in the slope of the rising curve, which indicates the beginning of the area of operation.

The new value for the operation voltage should be a bit above the position of the change in the slope. 

Figure 7 shows an example of a MCP curve with the specified operation voltage. The MCP curve for specifying
the starting operation voltage for a detector can always be found in the specification sheet.

Figure 7: Example of MCP curve (for illustration only).

The life time of MCPs can be very large, due to the high stability of the operation performance over a large
amount of extracted output charges after the burn-in period. The main reason for an often much shorter

                                                                                             17 

Summary of Contents for MCPD18

Page 1: ...MCPD18 R104 1 channel MCP Detector MCPD18 Release 104 Manual 1...

Page 2: ...r permission of Surface Concept GmbH Surface Concept GmbH Am S gewerk 23a 55124 Mainz Germany Tel 49 6131 627160 Fax 49 6131 6271629 www surface concept com support surface concept de User manual for...

Page 3: ...Operation of the MCPD18 8 4 1 Getting Started 8 4 1 1 Start Up Procedure 8 4 1 2 Dark Count Rate Measurement 9 4 1 3 Standard Detector Measurement 9 4 2 Standard Operation Procedure 10 4 3 Bake Out P...

Page 4: ...the microchannel plates and the MCP detector 2 2 Safety Instructions Please read this manual carefully before performing any electrical or electronic operations Strictly follow the safety rules given...

Page 5: ...l MCP detector It consist of a chevron microchannel plate stack and a detector anode MCP pulses are capacitively coupled into the anode The anode is a 50Ohm matched system and signal readout is realiz...

Page 6: ...ceed as follows to install it into your vacuum chamber Vent the package carefully Unpack the detector carefully Before installing the detector to your chamber check the front side of the MCP stack for...

Page 7: ...F for the high voltage supply and one BNC connector labeled SIG for the signal output MCP B connects the high voltage part of the detector anode and the back side of the MCP stack while MCP F connects...

Page 8: ...Ps very fast Turn off ion pumps and ion gauges before turning on the high voltage of the detector Turn on the high voltage carefully The voltage increase should not exceed 400V per minute A schematic...

Page 9: ...lary 1600V 4 1 2 Dark Count Rate Measurement Check the detector output by means of any electronics and software or with an oscilloscope where the input is terminated with 50Ohms The dark count rate wi...

Page 10: ...Observe the chamber pressure carefully every time the high voltage is turned on Switch off the high voltage immediately in case of a temporary pressure rise by an order of magnitude or more This indic...

Page 11: ...MCPD18 R104 Do not remove the blankets until the entire system has thoroughly cooled off Do not operate the detector before the temperature has returned to ambient conditions 11...

Page 12: ...l while the signal readout is realized close to ground potential in respect to the high voltage of the MCPs A pulse coupling layer is used to isolate the high voltage from the ground potential within...

Page 13: ...for the R104 The resistance between MCP F and MPC B resistance of MCP stack and internal resistor should be in the range of 100 400 M the exact values are given in the specification sheet of the detec...

Page 14: ...dling MCPs No physical object should come into contact with the active area of the wafer The MCP should be handled by its rims there is no solid glass border Use clean degassed tools fabricated from s...

Page 15: ...e maximum voltage given in P and in the specification sheet of the detector Higher potentials may result in irreversible damage MCPs can be degraded by exposure to various types of hydrocarbon materia...

Page 16: ...est count rates during the test phase Therefore gain degradation will be still significant in the first year of operation or even longer depending on applied count rates and becomes obvious by a decli...

Page 17: ...oltage Do as follows to measure an MCP curve Start the detector operation with a homogeneous illumination as good as possible Use an appropriate device to measure the count rate output of the detector...

Page 18: ...erated at a high count rate of several Mcps for a time period of about 8 hours It is really important that the detector is irradiated homogeneously over the complete active area Otherwise the MCPs wil...

Page 19: ...t detector see specification sheet Voltage at MCP front see specification sheet HV connections 2 SHV connectors MCP front MCP back Connections for signal 1 BNC connector 50Ohm matched coaxial signal M...

Page 20: ...tor elements 12 Figure 4 Internal connection of the high voltage potentials schematic for the R104 13 Figure 5 Typical gain degradation of MCPs as function of the extracted output charge 16 Figure 6 S...

Reviews: