21
SC-TDC-1100 Series Manual
SC-TDC-1100/02D - Release 312-50
• 2/12 NIM module
• Number of Stop Inputs: 2
• Number of Start Inputs: 1 (common start input usable as reset of the internal clock resolution adjust
mode: quartz-accurate, adjustable resolution, insensitive to temperature variations, adjustable via
software (no calibration necessary)
• Digital time bin resolution per channel: 27.4ps
• 5.5ns pulse-pair resolution on one channel and 0ns between two channels
• Trigger to rising edge
• Start retrigger rate (max.): 9MHz
• Measurement range: 0ns – 40µs in start-stop operation
(measurement range of 40µs corresponds to a start frequency of 25kHz)
• Extended measurement range of up to 27.4ps x 64bit (>5000d) in reference mode
• Dynamic range: 2E19
• All channels provide precisely an equal resolution
• 32-fold multi-hit capability per channel
• 40MHz internal device measurement rate
• Stop Signal Input: NIM on LEMO 00 socket
• External Start Signal Input: NIM on LEMO 00 socket
• Device Synchronization Signal Input: Low voltage TTL on 50Ohm BNC socket
• Device Synchronization Signal Output: Low voltage TTL on 50Ohm BNC socket
• USB Interface for Data Transfer
SC-TDC-1100/02D - Release 312-70
• 2/12 NIM module
• Number of Stop Inputs: 2
• Number of Start Inputs: 1 (common start input usable as reset of the internal clock resolution adjust
mode: quartz-accurate, adjustable resolution, insensitive to temperature variations, adjustable via
software (no calibration necessary)
• Digital time bin resolution per channel: 27.4ps
• 5.5ns pulse-pair resolution on one channel and 0ns between two channels
• Trigger to rising edge
• Start retrigger rate (max.): 9MHz
• Measurement range: 0ns – 40µs in start-stop operation
(measurement range of 40µs corresponds to a start frequency of 25kHz)
• Internal start frequency divider (2-, 4-, 8-, 16- and 32-fold divider)
• Dynamic range: 2E19
• All channels provide precisely an equal resolution
• 32-fold multi-hit capability per channel
SC-TDC-1100 Series Manual | Surface Concept GmbH