  How to Do Zero Adjust   

 Using the Zero Adjust button 

(1) Make sure that your Pressure Sensor is 

opened to the atmospheric state. 

(2)  Press and hold the Zero Adjust button for at 

least three seconds. 

 
 
 
 
 
-  The LED will flash. 

 
 
 
 
 
 
 
 

-  When you release your hand from the Zero 

Adjust button, zero adjust will start. 

 
 
 
 
 
 
(3)  After the zero adjust is complete, the LED 

lamp will change from a flashing state to a 
constantly lit state. The pressure sensor 
returns to measurement mode. 

 
[Zero adjust error] 

If the LED lamp keeps flashing after at least 
three seconds from the time zero adjust started, 
it is the indication of a zero adjust error. 

 

-  When zero adjust is in progress, an error 

will occur if the pressure in your piping 
system fluctuates by ±1% F.S. or more. 

 

-  To reset the error, press and hold the Zero 

Adjust button for at least one second. 

 

After the error resets, redo the zero adjust 
procedure from Step (1), as described in 

Using the Zero Adjust button. 

 

 

 Zero adjust by external input 

(1) Make sure that your Pressure Sensor is 

opened to the atmospheric state. 

(2)  Short circuit the Zero Adjust input signal for at 

least three seconds. 

 
 
 
 
 

-  When zero adjust starts, the zero adjust 

output alternates ON/OFF every 0.2 
seconds. 

-  Zero adjust begins when the Zero adjust 

input is opened. 

 
 
 
 
 
 
 
(3)  When zero adjustments is complete, the zero 

adjustment output transistor activate (short the 
output signal to 0V) and resume measurement   
mode. 

 
[Zero adjust error] 

If the zero adjust output keeps alternating 
ON/OFF after at least three seconds from the 
time zero adjust started, it is the indication of a 
zero adjust error. 

 

-  When zero adjust is in progress, an error 

will occur if the pressure in your piping 
system fluctuates by ±1% F.S. or more. 

 

- To reset the error, short circuit the zero 

adjust input for at least one second. 

 

After the error resets, redo the zero adjust 
procedure from Step (1), as described in 

Zero 

Adjust by external input. 

 
 

Zero adjust conditions: 

- Open to the atmospheric state. 
- Pressure fluctuation must be within ±1% F.S.

ZE

RO

US

D J

Zero adjust conditions: 

- Open to the atmospheric state. 
- Pressure fluctuation must be within ±1% F.S.

                         

CAUTION 

Before doing the zero 

adjust

 procedure, 

Pressure Sensor is opened

 to the 

atmospheric 

   

state. 

Before doing the zero 

adjust

 procedure, 

Pressure Sensor is opened

 to the 

atmospheric 

   

state.

                         

CAUTION 

Periodically provide the ZERO point 

adjustment. 

   

     

Summary of Contents for HPIX Series

Page 1: ...3UZX418145E 06 1 9 Fluorine Resin Type Pressure Sensor HPIX HPSX Operation Manual Surpass Industry Co Ltd 2020 5...

Page 2: ...possible liquid leakage The use of dangerous chemicals solvents and gases may cause physical impairment Do not apply more pressure than the allowed maximum output Doing so will cause product failure...

Page 3: ...the case for easy recognition of power on status lit and zero adjust flashing Specifications Model HPIX HPSX Type Inline type Straight type Standard Rated Pressure Range 0 to 500kPa see nameplate Wit...

Page 4: ...6 5 21 38 48 4 5 HPIX 14 20 47 58 21 38 48 4 5 HPIX 16 24 62 73 21 42 50 4 5 Cable HPIX Control unit Base Case Body Connector Control unit ZE HPSX Case Body Connector Cable LED lamp Zero Adjust button...

Page 5: ...re corrosive or explosive gases are being ejected mounting positions in which the connector part faces upward areas that may be struck by lightning areas that may be submerged in water When mounting c...

Page 6: ...t values Current 4 mA Voltage 1 V Externally displayed pressure 0 kPa Analog output values Current 4 mA Voltage 1 V Never do wiring work when power supply is on Doing so may cause electrical shock Do...

Page 7: ...Zero Adjust input signal for at least three seconds When zero adjust starts the zero adjust output alternates ON OFF every 0 2 seconds Zero adjust begins when the Zero adjust input is opened 3 When ze...

Page 8: ...by 1 F S or more Press and hold the zero adjust button for 1 or more second to reset the error Adjust the pressure in your piping so that it fluctuates less than 1 F S Then repeat zero adjust No anal...

Page 9: ...njury due to an unauthorized usage in combination with other products and an improper usage outside of the specifications Our product warranty shall be limited to replacement of product Replacement wi...

Reviews: