OPERATING
PROCEDURES
OP CIC1200 E Rev4.0 06-09 Chapter 4
25
4
4.8. Constant Intensity Mode
(CI Mode)
4.8.1. Set Light Intensity Level
(CI Mode)
Run the CIC in CP Mode (CP key is illuminated).
Select CH1 or CH2 by actuating the key CHANGE
DISPLAY and validate.
Then actuate the key SET CH1/CH2 for about 1-2
sec. The alphanumeric display LIGHT INTENSITY
shows:
the flashing value
given by the corresponding sensor
the display LAMP PWER shows
the rated operational power.
By actuating the keys
one can choose an appropriate value for the respec-
tive channel.
Note
:
The maximum output power of the 350W lamp is 400 [W]
The minimum output power of the 350W lamp is 200 [W]
(see Program Parameters of the CIC1200 for the various
types of exposure lamps, in table 3.1)
When reaching one of these max/min. values an
acoustical signal is activated to inform the oper-
ator!
Actuate the key SET CH1/CH2 for about 1-2 sec.
and store the values displayed. a high-pitched
„beep“ will be heard confirming the operator that the
parameters have been stored.
Now one can run the CIC1200 in Constant Intensity
Mode.
Switch subsequently to CHANNEL 1 (CH1) or
CHANNEL 2 (CH2) respectively (see Section
„Change Display“)
With software version < 2.00:
If the shutter of the lamp house is closed in the dis-
play of the CIC reads
IDLE XXXX
(the idle wattage of the lamp)
With software version from 2.00 you can read the
last used exposure wattage.
With shutter open the corresponding key - CH1 or
CH2 - is illuminated indicating the active state and
the display shows the value of the light intensity cor-
responding to your choice as indicated in the annun-
ciator group DISPLAY.
Closing the shutter again, the display reads:
IDLE XXXX
The operating conditions of the exposure lamp have
been set. The CIC runs the lamp according to the
present parameters. The alphanumeric display
shows the corresponding values for LIGHT INTEN-
SITY and LAMP POWER (power necessary to run
the lamp with the desired intensity).
+ -
Summary of Contents for CIC1200
Page 1: ...USER MANUAL Constant Intensity Controller CIC1200 Rev4 0 06 09 www suss com...
Page 2: ......
Page 4: ...DISCLAIMER AND PUBLISHING DETAILS II DISCLAIMER i...
Page 16: ...PROGRAM PARAMETERS 10 OP CIC1200E Rev4 0 06 09 Chapter 3 3...
Page 36: ...OPERATING PROCEDURES 28 OP CIC1200E Rev4 0 06 09 Chapter 4 4...
Page 40: ...MAINTENANCE AND SERVICE 32 OP CIC1200E Rev4 0 06 09 Chapter 5 5...
Page 41: ......