LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV.
32
Press the 1 key to start RUN the RING BUFFER. If you have failed to set a triggered
operational mode, you will see the following error message and the RING BUFFER will not
RUN.
If a “triggered” operational mode has been selected the following status message will be
displayed (MOVE ON SYNC, INHIBITED BY STROBE mode is used in this example).
To exit the buffer run, simply hold the LOCAL button down. The controller will not
recognize the key press unless it is active just after a filter change has been triggered.
3.7.9
Turbo-Blanking Mode
[MODE/2]
As the optical pathway is transitioning between two FILTER POSITIONS that are not
adjacent to one another, the Input Galvanometer must briefly direct the light through any
filter(s) intermediate to the desired FILTER POSITION. Under normal operation, the
Output Galvanometer moves in phase with the Input Galvanometer and, therefore,
maintains the optical pathway to the Light Guide Port as the filter change occurs. This
results in a brief output of light at an undesired wavelength prior to arriving at the selected
FILTER POSITION. This problem can usually be avoided by placing filters in adjacent
FILTER POSITIONS when it is known in advance that you will only want to switch between
those two wavelengths. This may not be a practical solution for several situations. The
TURBO-BLANKING operational mode eliminates the transmission of undesired
wavelengths during filter changes by forcing the Input and Output Galvanometers to move
out of phase with one another. In this way, the optical pathway to the Light Guide Port is
broken until the Input Galvanometer has arrived at the selected FILTER POSITION and the
Output Galvanometer catches up with it. This feature does increase the time required for
switching, though. This operational mode is selected by pressing the MODE key. The
following menu will be displayed:
To display the TURBO-BLANKING menu shown below, press 2.
ERROR! A CONTROL MODE MUST BE SELECTED
BEFORE YOU RUN THE RING BUFFER
FILTER SEQUENCE FROM RING BUFFER
MOVE ON SYNC/STROBE
0=EXIT, 1=NEXT PAGE, 2 = TURBO BLANKING
3 = DISPLAY THE FIRST 4 FILTER VALUES
Summary of Contents for Lambda DG-4Plus/USB
Page 3: ......
Page 4: ......
Page 16: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 4...
Page 46: ......
Page 70: ......
Page 76: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 64 This page intentionally blank...
Page 78: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 66 This page intentionally blank...
Page 94: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 82 NOTES...
Page 98: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 86 NOTES...