LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV.
38
Nominal Filter Pos. Alt-
key-
pad
entry
Ctrl
char
ASCII/
-ANSI
def./-
char.
Phys.
Filter
Pos.
Move On
…
Atte-
nua-
tion
Details
Dec. Hex Binary
30 1E 00011110 -0030 ^^
(Unused)
31 1F 00011111 -0031 ^_
(Unused)
NOTES:
1.
Nominal Filter Position (Bits 0 – 3): Value range 0 through 15, and with Bit 4 = 1, 16 through 31.
2.
Move on Command or Trigger (Bit 4): 0 (on command (i.e., immediately)) or 1 (by trigger (see Note 3)).
3.
“On trigger” means the light beam moves only on Strobe (STROBE TTL input).
Table 4-4. DG-5-Only Filter Position Movement Commands (no return data).
Nom. Filter Pos.
Alt-
key-
pad
entry
Ctrl
char
ASCII/
-ANSI
def./-
char.
Phys.
Filter
Pos.
Move On
…
Atte-
nua-
tion
Details
Dec. Hex
.
Binary
0 00 00000000 -0000 ^@
0 Command 0%
Move on command to non-filter pos. where light is blocked
(shutter closed)
1 01 00000001 -0001 ^A
1 Command
100%
Move on command to 1
st
filter (18mm) at full
2 02 00000010 -0002 ^B
2 Command
100%
Move on command to 2
nd
filter (18mm) at full
3 03 00000011 -0003 ^C
3 Command
100%
Move on command to 3
rd
filter (18mm) at full
4 04 00000100 -0004 ^D
4 Command
100%
Move on command to 4
th
filter (25mm) at full
5 05 00000101 -0005 ^E
5 Command
100%
Move on command to 5
th
filter (25mm) at full
6 06 00000110 -0006 ^F
1 Command
50%
Move on command to 1
st
filter (18mm) at half
7 07 00000111 -0007 ^G
2 Command
50%
Move on command to 2
nd
filter (18mm) at half
8 08 00001000 -0008 ^H
3 Command
50%
Move on command to 3
rd
filter (18mm) at half
9 09 00001001 -0009 ^I
4 Command
50%
Move on command to 4
th
filter (25mm) at half
10 0A 00001010 -0010 ^J
5 Command
50%
Move on command to 5
th
filter (25mm) at half
11 0B 00001011 -0011 ^K
1 Command
33%
Move on command to 1
st
filter (18mm) at 1/3
12 0C 00001100 -0012 ^L
2 Command
33%
Move on command to 2
nd
filter (18mm) at 1/3
13 0D 00001101 -0013 ^M
3 Command
33%
Move on command to 3
rd
filter (18mm) at 1/3
14 0E 00001110 -0014 ^N
4 Command
33%
Move on command to 4
th
filter (25mm) at 1/3
15 0F 00001111 -0015 ^O
5 Command
33%
Move on command to 5
th
filter (25mm) at 1/3
16 10 00010000 -0016 ^P
0 Trigger 0%
Move on command to non-filter pos. where light is blocked
(shutter closed)
17 11 00010001 -0017 ^Q
1 Trigger
100%
Move on trigger to 1
st
filter (18mm) at full
18 12 00010010 -0018 ^R
2 Trigger
100%
Move on trigger to 2
nd
filter (18mm) at full
19 13 00010011 -0019 ^S
3 Trigger
100%
Move on trigger to 3
rd
filter (18mm) at full
20 14 00010100 -0020 ^T
4 Trigger
100%
Move on trigger to 4
th
filter (25mm) at full
21 15 00010101 -0021 ^U
5 Trigger
100%
Move on trigger to 5
th
filter (25mm) at full
22 16 00010110 -0022 ^V
1 Trigger
50%
Move on trigger to 1
st
filter (18mm) at half
23 17 00010111 -0023 ^W
2 Trigger
50%
Move on trigger to 2
nd
filter (18mm) at half
24 18 00011000 -0024 ^X
3 Trigger
50%
Move on trigger to 3
rd
filter (18mm) at half
25 19 00011001 -0025 ^Y
4 Trigger
50%
Move on trigger to 4
th
filter (25mm) at half
26 1A 00011010 -0026 ^Z
5 Trigger
50%
Move on trigger to 5
th
filter (25mm) at half
27 1B 00011011 -0027 ^[
1 Trigger
33%
Move on trigger to 1
st
filter (18mm) at 1/3
28 1C 00011100 -0028 ^\
2 Trigger
33%
Move on trigger to 2
nd
filter (18mm) at 1/3
Summary of Contents for Lambda DG-4Plus/USB
Page 3: ......
Page 4: ......
Page 16: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 4...
Page 46: ......
Page 70: ......
Page 76: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 64 This page intentionally blank...
Page 78: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 66 This page intentionally blank...
Page 94: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 82 NOTES...
Page 98: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 86 NOTES...