TRIO MPC-145 SERIES MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.62C (20191001) (FW V2.62)
19
2.5.2.2
Switch 5: Y-Axis Lockout during Homing
Table 2-2. Configuration Switch 5: Configuring the Homing Y-Movement Lock Out.
Switch #
Definition
State
Setting Position
5
Y-Movement lock out
during homing
(pressing HOME or
WORK)
Enabled: Locked
(no movement while X & Z are homing)
OFF
UP
Disabled: Not locked
(Y moves along with X & Z when homing)
ON*
DOWN*
* Factory default (recommended normal operation setting)
2.5.2.3
Switch 6: Calibration Homing on Power On
Table 2-3. Configuration Switch 6: Configuring the Sensor Test.
Switch #
Definition
State
Setting Position
6
Calibration Homing
on Power On
Disabled: No calibration occurs on power on.
Power-off position is retained on power on
OFF
UP
Enabled: Calibrates to 1,000 µm for all axes on
power on. Power-off position is forgotten.
ON*
DOWN*
* Factory default (recommended normal operation setting)
2.5.2.4
Switches 7 & 8 Reserved
Switches 7 & 8 are reserved.
Table 2-4. Configuration Switches 7 & 8: Reserved.
Switch #
Definition
Setting Position
7
Reserved
OFF*
UP*
8
Reserved
OFF*
UP*
* Factory default (recommended normal operation setting).
2.5.2.5
Switch 9 Electromechanical Device Compatibility
Table 2-5. Configuration Switch 9: Electromechanical device compatibility.
Switch #
Definition
State
Setting Position
9
Electromechanical
device compatibility
MP-845[S]/M, MP-245/M micromanipulators
OFF*
UP*
MP-285/M micromanipulator; 3DMS & MT-78
stages; MOM & SOM objective movers
ON
DOWN
* Factory default (recommended normal operation setting).