background image

Should your SVANTEK professional measurement equipment need to 
be returned for repair or for calibration, please contact the service 
office at the following number or contact via the SVANTEK’s website.

Service Office: +48 (22) 51-88-320 or +48 (22) 51-88-322.
Office hours are 9:00 a.m. to 5:00 p.m. Central European Time.

-E-mail at office@svantek.com  

-Internet at www.svantek.com 
-Address: 
 

SVANTEK Sp. z o.o.

 

Strzygłowska 81

 

04-872 Warszawa, 

 Poland

SV 103 User Guide

8

Support and service

Summary of Contents for SV 103

Page 1: ...SV103 Hand Arm Vibration Dosimeter User Guide ...

Page 2: ...107 moulded hand held adaptor with the tri axial MEMS accelerometer and contact force sensors The MEMS technology means the SV 107 is extremely robust shock resistant uses very low power and is free of the DC shift effect that adversely affects systems based on piezoelectric accelerometers ISO 5349 2 and EU Directive 2002 44 EC List of Contents Page 3 CONTENTS OF THE SV 103 BOX Page 4 CONFIGURATIO...

Page 3: ...es sending the instrument to SVANTEK authorised laboratory SV_CV_HA_T Accredited calibration for vibration meter ISO IEC 17025 calibration certificate List of contents included in the SV 103 box Please check the list against the contents If any of the listed items are missing please contact your local Svantek distributor or send an e mail to office svantek com within 3 days of receiving the parcel...

Page 4: ...3 settings using panels on the right hand side 8 Disconnect the USB cable Instrument is ready to measure SV 103 User Guide 4 2 Connect the USB cable to PC Make sure the Supervisor software and USB drivers are previously installed on the PC 6 Save settings file under new name 7 Upload the file to the Instrument Files using left arrow and select Apply 3 Open the Supervisor software and go to the Ins...

Page 5: ...calibrator 5 Start the calibrator using level of 10 m s2 at 80 Hz and wait until all LEDs are green 6 Press Enter on SV 103 to start the check procedure 7 Repeat the procedure for all 3 axes SV 103 User Guide 5 4 Enable Calibration menu in SV 103 by holding Right Arrow button Select the System check and axis SV110 calibrator and SA105A adapter are not the part of the SV103 kit They need to be orde...

Page 6: ...ry to use a screw for measurements Set up of SV 103 5 To start the measurement manually pres 2 arrows simultaneously SV 103 User Guide 6 4 Place the SV 107 on the palm of the hand as shown below Secure the cable on a wrist with a wrist band 2 Put the mounting adapter on arm and place the SV 103 in it ...

Page 7: ...V 103 User Guide 7 2 Select files you wish to download and click the Download button located in the bottom section 3 Once the file is downloaded it will be automatically shown in the Data Browser Right click on file and start New session or simply double click on file 5 The detailed information on the operations within the Session panels is described in the SV 103 User Manual and Supervisor Softwa...

Page 8: ...fice at the following number or contact via the SVANTEK s website Service Office 48 22 51 88 320 or 48 22 51 88 322 Office hours are 9 00 a m to 5 00 p m Central European Time E mail at office svantek com Internet at www svantek com Address SVANTEK Sp z o o Strzygłowska 81 04 872 Warszawa Poland SV 103 User Guide 8 Support and service ...

Page 9: ...1 dB Measurement Range 0 2 ms 2 RMS 2000 ms 2 PEAK Frequency Range 1 Hz 2000 Hz Data Logger Time history data including meter mode results and spectra Time Domain Recording Simultaneous 3 channel time domain signal recording option Analyser 1 1 octave real time analysis option 1 3 octave real time analysis option Accelerometer Detachable SV 107 tri axial accelerometer with hand straps in accordanc...

Reviews: