39
4.14
Reviewing measurements
Most of the parameters can be inspected in real-time during either measurement run or stopped mode
of operation. If the display screen is turned off just press any key (but watch notice below).
The keys on the keypad of the instrument enable you to navigate thru most of the parameters.
For specific information on the VIEW modes, please see chapter 3.9.
Use the
<SCROLL>
key to move down the list through the various measurements
Use the
<PROFILE>
key to change the ACOUSTIC PROFILE you would like to inspect
Use the
<ENTER>
key to change the VIEW mode
Notice:
In most cases the keypad will be probably locked. To gain access to the results and
unlock the keypad see chapter 4.12.
Notice:
After reviewing results remember to lock the keypad again in order to maintain
the integrity of the measurement run by preventing uncontrolled access to the instrument.
4.15
Resetting the dosimeter
SYSTEM RESET:
(internal software reset clears any setup configuration, and brings back the
default factory settings).
See
Appendix A
for remote control commands description.
HARDWARE RESET:
(internal hardware reset, no user data is changed)
Make sure the battery is not exhausted, and the unit is turned off. Hold down the
<ENTER>
key
for more than 15 seconds, and then release it. Turn on the instrument as usually (see chapter 4.3).
Notice:
Hardware reset is only to be used in extreme situations such as an instrument hang-
up.
Be aware, that a hardware reset:
- will stop any pre-programmed auto-run modes
- will stop measurement run
- HARDWARE RESET works, even if the keyboard becomes locked out!