Transportation and carrying
For transportation or storage purpose, always use the packaging provided by the manufacturer. In a
potentially dirty industrial environment it is advisable to use the carrying case provided by the manufacturer
such as the
SA 73
(see chapter 2.3), which ensures excellent mechanical and environmental protection and
long term storage conditions.
6.9
Troubleshooting
1. Upon connection to the USB port, if automatic charging is not started: check the USB cable and
power supply ratings of the source.
2. If the incorrect time or date is displayed when turning on the instrument connect the device to the
computer and use Supervisor software to set the time and date (see) ensuring PC clock is set
correctly
3. In case the instrument is not able to turn on ensure the unit is charged by connecting to USB
or recommended charger. This ensures the battery is not exhausted. Then proceed with hardware
reset (chapter 4.15)
4. In case your dosimeter does not respond proceed with turn-off/turn-on procedure (chapter4.3),
and hardware reset of the instrument (chapter 4.15).
5. In case the measurement of the sound level is frozen or set to a fixed value proceed with turn-
off/turn-on procedure (chapter 4.3), then with hardware reset of the instrument (chapter 4.15).
6. In case the reset does not help proceed to chapter 7.