113
SV 200A User Manual
CLD
window displays list of stations running in the
CLD
mode along with the current state of each
station. The progress bar indicates activity of the
CLD
mode.
Configuration
To configure
Continuous Logger Download
session settings, click
Options
button. Continuous logger
download options window will
appear allowing you to make necessary adjustments.
Options
button is only active when
CLD
is not running.
Download step
determines how often data from SV 200A
will be downloaded to the PC.
File saving settings
enable configuring basic settings for files download such as download path, file
naming rules and writing session log.
Daily data settings
enable configuring the path where files downloaded from the instrument will be
copied after each day of measurements. Files may be automatically merged, if more than one file is
generated throughout the day.
Archivisation settings
enable configuring the path where all downloaded files are copied for
archivisation purposes. Files may be compressed into monthly zip file.
The
System validation & summary
section enables configuring the following settings:
•
time of station summary (hourly, daily, weekly or monthly)
•
clearing downloaded loggers from the instrument during measurements
•
system checks with built-in electrostatic actuator,
•
automatic synchronization of real time clock,
•
verification of instrument settings during station summary.
The
Publish WWW/SVN Upload
section enables configuring basic
settings for HTML publishing of the results and FTP uploads. Detailed
settings can be configured for each individual station by clicking WWW
button which becomes visible in the station list.
The
Configure FTP
button enables configuring the FTP server which is going to be used for data upload.
download
step
daily data
settings
archivisation
settings
system
validation &
summary
file saving settings
publishing
settings