115
SV 200A User Manual
Note: More information on Live Results can be found in the + User Manual
available online:
http://svantek.com/lang-en/support/19/svanpc_software.html#PDF.
Basic information
The
Live Results
mode runs as a Windows service. It does not require + application to be
running. Windows user who runs the LR does not have to be logged on the PC. LR
automatically
resumes to its previous state after booting the PC.
Live Results
window displays list of stations running in LR
mode along with the results selected for
display.
Configuration
You can configure results that are to be displayed for each station separately. The station currently
being configured can be selected in the
Station
combo-box. Results are displayed with seven possible
time intervals: 1 s, 2 s, 5 s, 10 s, 20 s, 30 and 60 s. This setting is controlled by
Read step
combo-box.
Read step
setting can be altered regardless of the
Live Results
session
being started.
With the
Live Results
mode stopped you can select particular results for displaying after clicking the
Configure results
button.
The
HTML publishing
section enables configuring the path of HTML and CSV files containing current
measurement results.
Starting session
To start the
Live Results
session click
Start
button. The LR
session will start immediately.
Measurement status control
While the
Live Results
session is active the
Start/stop
measurement
button allows you to start or stop measurement
manually.
Actuator control
While the
Live Results
session is active the
Enable/disable
actuator
button allows you to control the state of the built-in
electrostatic actuator manually. After the actuator has been
enabled you can read the result thereby conducting manual
system check.
Stopping session
To finish
Live Results
click
the
Stop
button. The session will end
immediately.