17
INSULATION MODE
Measuring hookup:
Measuring section:
Menu:
K factor – to search the faults.
Threshold MOhm – setup for the bottom indicator - line (imitating the pointer
movement). It is convenient for the rapid check of cable pair good condition.
U test – test voltage: 400 V for defects searching, 180 V for work without turning off
the equipment.
Averaging – quantity of
fi
lters averaging: during the unstable indications it is
recommended to increase the averaging quantity. For rapid measurements you should
to decrease the averaging quantity.
Filter: auto – regular restart of the
fi
lter, manual – start by the [
ОК
] key. Before
turning on the manual
fi
lter it is necessary the cables complete charging. The
fi
lter
work is accompanied by bottom indicator movement.