AMI Solicon4
Installation
A-96.250.321 / 191015
19
3.2.
Mounting of Instrument Panel
To mount the instrument panel proceed according the following list:
The instrument must only be installed by trained personnel.
Mount the instrument in vertical position.
For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation
material is available:
– 6 Screws 6x60 mm
– 6 Dowels
– 6 Washers 6.4/12 mm
Mounting
requirements
The instrument is only intended for indoor installation, dimensions
see
.
3.3.
Connecting Sample Inlet and Outlet
Sample inlet
Use a plastic tube (FEP, PA, or PE 10x12 mm) to connect the sam-
ple inlet at the flow cell.
If a deltaT senor is installed, push the plastic tube over the hose
nozzle of the deltaT sensor.
Sample outlet
Connect the flexible tube 20x15 mm [C] to the 1/2” hose nozzle [B]
and place it into a drain with atmospheric pressure.
A
B
Plastic tube 10x12
Elbow hose nozzle
A
B
A
B
C
Waste funnel
1/2” Hose nozzle
Flexible tube 20x15 mm
A
B
C