AMI Solicon4
Program List and Explanations
A-96.250.321 / 191015
53
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5
Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Message List
1.2.1
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation
:
Designation of the instrument.
Version
: Firmware of instrument (e.g. V6.00-10/15)
2.1.3
Factory Test
: Test date of the Instrument, Motherboard and Fron-
tend
2.1.4
Operating Time:
Years / Days / Hours / Minutes / Seconds
2.2 Sensors
2.2.1
Cond. Sensor
Current value
in µS
Raw value
in µS
Cell Constant
Contamination
2.2.1.5
Cal. History
: In this menu the calibration values of the last
calibrations are saved.
Number: Numbers the calibrations in descending order.
Date, Time: Date and time of each calibration.
Cell constant: Shows the cell constant of the sensor in use.
Max. 64 data records are memorized. One process calibration cor-
responds to one data record.