AMI Solicon4
Program List and Explanations
A-96.250.321 / 191015
57
5 Installation
5.1 Sensors
5.1.1
Sensor parameters
5.1.1.1
Cell Constant
: Enter the cell constant printed on the sensor label.
5.1.1.2
Temp. Corr
: Enter the temperature correction printed on the sensor
label.
5.1.1.3
Cable length
: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
5.1.2
Temp. Compensation
5.1.2.1
Comp
.: Available compensation models are:
none
Coefficient
non-linear DIN
5.1.3
Flow
5.1.3.1
Flow measurement
: Select the type of flow sensor if a flow sensor
is installed.
Possible flow sensors: None; Q-Flow; deltaT.
5.1.3.2
Slope
: If flow measurement is set to deltaT.
The slope value is used to adjust the flow measurement of the
deltaT sensor if the ambient temperature is higher or lower
than 20 °C.
5.1.4
Conc.
The menu <Concentration> (Conc.) allows the additional measure-
ment of a known substance in the sample. The concentration of the
substance is calculated based on the conductivity of any of the fol-
lowing substances. The calculated value is displayed in %. As an
exception, TDS is displayed in mg/l.
none
nitric acid
hydrochloric acid
sodium chloride
caustic soda
sulfuric acid
salinity
TDS as NaCl
TDS