background image

 

6

3.  While viewing through the eyepiece, rotate the coarse focusing knob 

to bring the specimen into view.    Next, sharpen the focus using the 
fine focus knob.

 

4.  If the image of the specimen appears pale, the aperture of the 

diaphragm should be decreased (if the model has a disc diaphragm, 
rotate the disc to a smaller hole, conversely, if the model has the iris 
diaphragm, it should be closed slightly).  This will increase contrast 
in the specimen’s image.    If the specimen appears dark, slightly 
open the diaphragm.     

5.  Rotate the nosepiece to the next higher  power objective. A slight 

turn of the fine focusing knob may be required to bring the image of 
the specimen into sharp focus. Once the specimen is in focus with the 
highest power objective, it will be in focus with each lower power 
objective. 

 
 

OIL IMMERSION (Only for models with a 100X objective) 

 

When light rays from the objective lens to the specimen pass through air, 
they are distorted slightly, a phenomenon known as refraction. This is 
usually not a problem at a magnification of 400X or lower. However, at a 
magnification of 1,000X and above, refraction becomes problematic.   
This problem is reduced significantly by placing a thin layer of very clear, 
viscous oil between the slide and tip of the objective lens. The result is a 
much clearer image at 1,000X because the oil has the same light 
transmitting properties as glass. Using oil slightly increases the 
resolution and brightness of the image. Usually a very thin slide (size #1) 
is used for oil immersion because at this magnification, the working 
distance is very small and is critical to focusing the specimen. Good 
quality glass (not plastic) cover slips should be used. If their thickness is 
over 0.17mm, the objective will not resolve properly, because the 
specimen cannot be moved close enough to the objective lens to be in 
focus. 
 
Place a tiny amount of oil (only 1 drop should be sufficient) onto the slide 
prior to rotating the 100XRD objective into position. It is essential to 
thoroughly clean the objective tip after use to prevent damage and to 
ensure that an image can be seen clearly the next time the objective is 
used. Please contact Swift Optical or your authorized Swift dealer for the 
appropriate immersion oil to use.     

Summary of Contents for M3500 SERIES

Page 1: ...CROSCOPE The Swift M3500 Series microscope is considered to be the most Student proof microscope on the market It is an instrument of optical and mechanical precision and will perform satisfactorily w...

Page 2: ...age drift CONDENSER the condenser is mounted in the stage and it is used in conjunction with the iris diaphragm The function of the condenser is to provide full illumination to the specimen plane and...

Page 3: ...ide in place allowing the user to move the slide on any x y axis through the manipulation of two control knobs or a co axial control mechanism NOSEPIECE the revolving turret that holds the objective l...

Page 4: ...COVER GLASS Thin glass cut in circles rectangles or squares for covering the specimen usually a thickness of 0 15 to 0 I7mm The majority of specimens should be protected by a cover glass and must be c...

Page 5: ...esolving power is directly proportional to the numerical aperture of the optical system WORKING DISTANCE the distance from the lens of the objective to the cover slip on the slide when the specimen is...

Page 6: ...ification of 400X or lower However at a magnification of 1 000X and above refraction becomes problematic This problem is reduced significantly by placing a thin layer of very clear viscous oil between...

Page 7: ...100XRD objective requires an iris diaphragm for brightfield oil immersion microscopy HOW TO USE THE POINTMASTER EYEPIECE RETICLE Swift s patented POINTMASTER eyepiece reticle which is installed in the...

Page 8: ...ware must first be installed on a computer The minimum computer requirements to use the camera is having an available USB 2 0 port Windows XP Vista or Mac OS X operating system installed on the comput...

Page 9: ...e microscope is not in use CLEANING The front lens of the objectives particularly the 40XRD and 100XRD should be cleaned after use First brush with a soft camel hair brush or blown off with clean oil...

Page 10: ...ay corrode your microscope Also be sure that the objectives are placed in the lowest possible position and the rotating head is turned inward and not protruding from the base Microscopes with mechanic...

Page 11: ...s electrical cord from the electrical socket and remove any slides on the stage Use the small allen wrench 09mm that was included with the microscope to loosen the set screws that hold the black illu...

Page 12: ...e enough Slowly open the diaphragm to increase the size of the opening allowing for additional illumination E PROBLEM Once the specimen is in focus it moves out of focus CORRECTION Gravity is causing...

Page 13: ......

Reviews: