background image

 

USING YOUR SWIFT MICROSCOPE 

 

1.  Use the stage clips or slide holder mechanism to secure the slide in 

place. Be sure the specimen is centered over the opening in the 
stage. 

 
2.  Rotate the nosepiece to place the lowest power (4X) objective over 

the specimen. Be sure the objective “clicks” into position. 

 
3.   Turn on the illumination by pressing the on/off switch towards “1” 
 
4.   (M3700 series only) Adjust the Siedentopf binocular head (by moving    
      the eyepiece tubes up and down in an arc-like motion, similar to  
      adjusting binoculars) until one perfect circle is seen in the field of  
      view. 

 

5.   Open the iris diaphragm to its largest aperature. 

 

6.   While viewing through the eyepiece(s), rotate the coarse focus knob   
      slowly and carefully to bring the specimen into focus. The specimen  
      may require some centering in the field of view at this time. By   
      using the fine focusing knob, slowly and carefully refine the focus to  
      clearly observe the fine details of the specimen. 
  
7.  If the image of the specimen appears pale, the aperture of the iris 

diaphragm should be slowly closed until the details of the specimen 
are sharply defined. If the specimen appears dark, slightly open the 
diaphragm. 

Please note: a smaller iris diaphragm aperture (opening) increases the 

contrast in the image while a larger aperture decreases the contrast. 
(The diaphragm is not intended for controlling the brightness of the 
illumination). A good procedure to follow in selecting the proper 
opening is to start with a large aperature and reducing it until the fine 
detail of the specimen is in exact focus. Using an inappropriate aperture 

results in a “washing out” of the image. Care must be exercised not to 
reduce the aperture too much to gain high contrast, as then the fine 
structure in the image of the

 

specimen will be destroyed. Reducing the 

aperture increases contrast and depth of focus, but it also reduces 
resolution and causes diffraction. Example: The aperture for the 10X 
objective will not be the same as for the 40XRD objective, since the 
angle of the required light is determined by the numerical aperture 
(N.A.) of the objective. The proper aperture of the diaphragm can be 
easily achieved after minimal experience with the microscope.

 

Summary of Contents for M3600 series

Page 1: ...Swift M3600 M3700 Series Microscope Use and Care Manual SWIFT OPTICAL Enduring Quality and Technical Excellence...

Page 2: ...nob Illuminator Rheostat Diopter Adjustment Arm Nosepiece Base Eyepiece Iris Diaphragm Illuminator M3600 Head Cord Holder Coarse Focus Knob Slide Holder Mechanical Stage Controls Fine Focus Knob Coars...

Page 3: ...located on both sides of the microscope The control is designed for a continuous operation over the range of the stage movement CONDENSER the condenser is mounted in the stage and it is used in conjun...

Page 4: ...denser allowing the user to control contrast MECHANICAL STAGE an alternative to stage clips is a mechanical stage A mechanical stage holds the slide in place allowing the user to move the slide on any...

Page 5: ...onduct light amplify magnification and convert the image into a field of view easily seen by the human eye COVER GLASS thin glass cut in circles rectangles or squares for covering the specimen usually...

Page 6: ...hat only a slight turn of the fine adjustment is required when a change is made from a lower to higher power RESOLVING POWER the ability of a lens to clearly separate fine detail Resolving power is di...

Page 7: ...aperture of the iris diaphragm should be slowly closed until the details of the specimen are sharply defined If the specimen appears dark slightly open the diaphragm Please note a smaller iris diaphra...

Page 8: ...agnification of 1 000X and above refraction becomes problematic This problem is reduced significantly by placing a thin layer of very clear viscous oil between the slide and tip of the objective lens...

Page 9: ...n microscopy HOW TO USE THE POINTMASTER EYEPIECE RETICLE Swift s patented POINTMASTER eyepiece reticle which is installed in the eyepiece of the microscope enables the user to easily measure the size...

Page 10: ...nstalled on the computer 1GB of RAM 1GB free hard drive space and 2Ghz CPU Complete instructions on how to use the software is included on the software CD that was packaged with the M3602C 3DGL and M3...

Page 11: ...e to create a small box in an area of the image that should be displayed as a white color The size of the box does not make a difference as long as it is only in an area that should be displayed as wh...

Page 12: ...echnician DUST COVER AND STORAGE All microscopes should be protected from dust by a dust cover when in storage or not in use A dust cover is the most cost effective microscope insurance you can buy En...

Page 13: ...h the holes at the bottom of the new LED and push the LED onto the socket Re install the illuminator housing COMMON PROBLEMS IN MICROSCOPY CAUTION Never disassemble mechanical or optical components Th...

Page 14: ...The focusing tension should be increased to act as a brake to prevent this downward drift The M3600 Series is outfitted with a tension collar on the course focusing shaft to allow tension on the slip...

Page 15: ...warranty is at the technician s discretion Other than set forth above Swift hereby disclaims all warranties expressed or implied of fitness for a particular purpose Defective products covered by the w...

Reviews: