background image

Issue 

Possible cause 

Solution 

Double images 

Interpupillary distance is 
not correct 

Readjust the eyepieces 

Diopter adjustment is not 
correct 

Readjust the diopter rings 

Magnification of each 
eyepiece is not the same 

Mount eyepieces of the 
same magnification 

Dirt appears in the field 
of view 

Dirt on the specimen 

Clean the specimen 

Dirt on the surfaces of 
eyepiece 

Clean the surfaces 

Image is blurry 

Objective surfaces are 
dirty 

Clean the objective 

Diopter adjustment is not 
correct 

Readjust the diopter 

Focus is incorrect 

Readjust the focus 

The focusing knob is not 
smooth 

The focusing knob is too 
tight 

Loosen it to a suitable 
position 

Head keeps slipping 
down 

The focusing knob is too 
loose 

Tighten it to a suitable 
position 

Eyes fatigue easily 

Diopter adjustment is not 
correct 

Adjust the diopter rings 

Illumination does not 
work when the switch is 
on 

No power supply 

Check the connection 
with the power supply 

 

 

Summary of Contents for S41-20

Page 1: ...User Guide the Swift S41 20 Stereo Microscope...

Page 2: ...help you begin using it For more information about our microscopes software downloads product manuals or customer service requests please visit www swiftoptical com Table of Contents I Specifications...

Page 3: ...eces WF10X WF20X Objective 1X Total magnifications 10x 20x Working distance 230mm Focusing Focus travel range 41mm Max specimen thickness 67mm Stand Single arm stand 18mm vertical pillar vertical pill...

Page 4: ...entified Remove the plastic covers from the microscope s head and the two gooseneck LED lights Assemble the stand connector and its supporting pillar on the vertical pillar tightening the fastening kn...

Page 5: ...hin the travel range of the S303 S304 pole arm and stand adjust the microscope head height along the vertical pillar to a working distance of 230mm from the sample Remember to tighten the locking scre...

Page 6: ...cover the microscope with the dust cover provided and store it in a dry and clean place free from moisture to prevent rust Remove the eyepieces when not in use and place covers on the eyepiece holder...

Page 7: ...yepiece Clean the surfaces Image is blurry Objective surfaces are dirty Clean the objective Diopter adjustment is not correct Readjust the diopter Focus is incorrect Readjust the focus The focusing kn...

Page 8: ...rom our warehouse for warranty services The warranty does not cover damages resulting from normal wear and tear abuse or unauthorized repairs For customers living outside the United States Swift Optic...

Page 9: ...improvements occasionally involve small modifications to the mechanical structure and optical design of our microscopes Therefore some descriptions illustrations and specifications in this instructio...