QSG Diode Pointer  Ver 2.1

a Novanta company

Diode Pointer Quick Start Guide

Coarse Adjustment:

14.  

Assure the laser is off! 

Loosen the four diode pointer mounting hex head screws on the face of the  

 

laser enough to move the pointer in the direction necessary to center the beam in the CO

2

 burn.   

 

Re-tighten and repeat the far and near field adjustment steps. If this does not bring the pointer into  

 

adequate coincidence with the laser, proceed to the next step.

 IF further coarse adjustment needed:

 Loosen  

  the screws holding mounting plate to pointer.

 Rotate the pointer mounting plate  

  180 degrees

: re-tighten the screws  

  holding mounting plate to pointer.

 

Re-install and tighten the two screws.

If the pointer is all the way to its limit in the slots, it may be necessary to rotate the 

mounting plate in relation to the pointer housing. 

15.

 

Assure the laser is off! 

Remove the diode pointer from the laser. 

16.

  Refer to the figures below and remove the two socket head cap screws holding the mounting plate to    

 

the pointer body. Rotate the mounting plate 180 degrees then reinstall and tighten the two screws.  

 

 

Reinstall the pointer to the front of the laser and repeat the near/far field adjustment as you did before    

 

until the pointer is in adequate coincidence with the laser beam.

Important

Note:

Summary of Contents for Diode Pointer

Page 1: ... following sections for guidance If your SYNRAD laser is shipped with a Diode Pointer installed it is also factory pre aligned and ready for operation 2 Upon arrival inspect all shipping containers for signs of damage If you discover shipping damage document the damage photographically if possible then immediately notify the shipping carrier responsible party for any transportation damage and SYNR...

Page 2: ... near far field adjuster screws accessible Tighten the four screws with the ignition wrench 4 1 Hanging the Diode Pointer onto the laser faceplate providing power to the unit Warning Serious personal injury Do not stare directly into the red laser light emitted from the Diode Pointer OR the laser The light is intense enough to injure the eye with exposure Always wear eye protection CO2 laser emits...

Page 3: ... on the cable connect the cable by the 5 VDC jack reinstall the hex nut on the outside of the faceplate affixing the cable Re install the face plate and refer to the prior page for mounting 6 2 Locate the laser s 5VDC jack shipped with your unit Remove the hex nut on the cable and set it aside 6 3 Insert the cable through the face plate power port and re install the hex nut 6 4 Connect the cable t...

Page 4: ...C 100Ma On Off Power supply 7 If your laser does not provide a power connection for the Diode Pointer use the plug from the supplied wall plug transformer into the 5 VDC jack on the side of the pointer 8 Press the On Off switch The red LED below the switch illuminates to indicate the pointer is on and laser diode emission should be visible ...

Page 5: ... 20 watts Note the lasers coincidence with the diode pointer Near field distance adjustment screws moves the beam vertical and diagonal A Near field Approx 30cm 30cm 9 1 Example of laser and diode pointer coinci dence within 0 1mm Diode Pointer install and setup complete 9 1 Example of a diode pointer that requires adjustment 10 If the beam requires adjusting insert the two Allen wrenches included...

Page 6: ...did in the prior steps Power on and fire the CO2 laser again at low power approxi mately 10 20 watts until you get a good burn image on the thermal card and observe the diode point ers coincidence with the beam You should have good coincidence by now If not see the above Im portant note before proceeding to the next section B Far field Approx 3 10m Important Note Since adjusting the beam affects b...

Page 7: ...inter Rotate the pointer mounting plate 180 degrees re tighten the screws holding mounting plate to pointer Re install and tighten the two screws If the pointer is all the way to its limit in the slots it may be necessary to rotate the mounting plate in relation to the pointer housing 15 Assure the laser is off Remove the diode pointer from the laser 16 Refer to the figures below and remove the tw...

Page 8: ... erance of alignment Anything more precise and accurate is desirable but not necessary Final accuracy and precision alignment is completely dependent on the final use of the pointer Most articulated arm applica tions require 100 micro radians while a simple lens system requires only 1 milliradian The outside tolerance or acceptable misalignment is 1 milliradian or 1 mm at 1 meter Congratulations T...

Reviews: