getting started
1
10
Synrad
Firestar i401
operator’s manual
Mounting
3
Carefully place the i401 laser on the mounting surface so the M10 threaded holes in the baseplate
line up over the thru holes in the mounting surface.
Important Note:
Verify the correct fastener length for your mounting application. The M10 × 1.5 ×
35 mm capscrews the ship kit are for use with the factory-installed mounting feet.
When fastening the i401 to your mounting surface from the bottom up, use M10 ×
1.5 mounting screws with a
length of 30 mm +/- 2 mm plus the thickness of the
mounting plate and any washers between the bolt head and the mounting plate.
4
Place a split washer and flat washer on each capscrew and insert the fasteners through the mounting
surface into the laser baseplate. Turn the screws by hand until the threads fully engage.
5
Evenly tighten all four fasteners to a torque of 40 N m (29 ft lb
f
).