Illuminator Replacement
As with any intense illumination source, the v300/v600 Illuminator Module will wear
out with use. High output cooling of the Illuminator Module extends the expected
life to 400+ hours under normal conditions. Spare Illuminator Modules can be
purchased to enable the user to continue use with minimal interruption in the event
of an Illuminator Module failure. It is highly recommended that a spare Illuminator
Module be kept on hand.
Please use the following procedures to replace the Illuminator Module in either your
v300 or v600 Visualization System. If you have any questions, please contact our
service group at 800.714.1374 or 207.657.7050.
Illumination Condition Indicator
The v300/v600 Series is equipped with one of the most sophisticated diagnostic
systems of any lighting or vision system in the medical world. A powerful
microprocessor constantly monitors the entire system of the lighting and cooling to
ensure maximum performance. Part of this computer’s function set is a display of the
Illuminator Module’s (1 in Fig 1) condition. Three colored LED’s are installed to be
visible through the air outlet vents on the rear of the illuminator.
Green:
Will remain visible for the first 75% of its life expectancy.
Yellow:
Will indicate 5-24% of life expectancy remains. It is suggested to order
replacement at this point.
Red:
Will indicate less than 5% of life expectancy remains. The bulb will alternate
dim and bright 7 times at startup during this critical stage showing that
the illuminator module has reached its life expectancy.
The LED’s flash a readout of the elapsed hours on the Illuminator Module 30 seconds
after the power is applied. The red LED flashes the number of hundreds of hours,
the yellow flashes the number of 10s of hours and the green flashes the number of
single hours, so 124 hours would be 1 red, 2 yellow and 4 green. This is followed
in another 15 seconds by a second series of flashes showing the number of on/off
cycles. This sequence differs from the first in that it is divided by 10, so 120 would be
1 yellow and 2 green flashes.
A typical life span for an Illuminator Module (1 in Fig 1) is in excess of 400
hours. Should your illuminator fail prior to 400 hours, a pro-rated warranty
will apply.
Summary of Contents for v300TM
Page 2: ...Revision 3 July 2004 English v300TM v600TM Visualization System Owner s Manual ...
Page 19: ...18 www syrisscientific com Syris Scientific LLC ...
Page 21: ......
Page 22: ...Revisie 3 juli 2004 Dutch v300TM v600TM Visualisatiesysteem Gebruikershandleiding ...
Page 41: ......
Page 42: ...Système de visualisation v300TM v600TM Révision 3 Juillet 2004 French Manuel d utilisation ...
Page 61: ......
Page 62: ...Revision 3 Juli 2004 German Visualisierungssystem v300TM v600TM Benutzerhandbuch ...
Page 81: ......
Page 99: ...18 www syrisscientific com Syris Scientific LLC ...
Page 101: ......
Page 102: ...Revisión 3 Julio 2004 Spanish Sistema de visualización v300TM v600TM Manual del propietario ...
Page 121: ......