Trouble-shooting
If lifting motion cannot be activated, check to ensure:
- that the emergency stop is not activated; on certain models, the emergency stop symbol lights up;
- that the lift unit is switched on (ON) via the hand control (a signal is heard);
- that the hand control is properly connected and pressed in securely; disconnect, and then replace it and press it in
securely;
- that the lift unit is not in the charging position; the lift unit is blocked for use when it is in the charging position;
- that the battery is charged;
- that the charger has a power supply; the charging lamp on the charger unit should light up when the lift unit is placed
in the charging position;
- that load is applied to the lift strap.
If the lift unit is not working properly, contact your Handicare dealer.
If the lift cannot be run all the way up to stop, check to ensure:
- that the overload symbol is not activated;
- that the battery symbol does not indicate low battery charge level.
If the lift cannot be run down, check to ensure:
- that load is applied to the lift strap;
- that the battery symbol does not indicate low battery charge level;
- that the directional indicators on the buttons, arrow up and arrow down, are correct.
If unusual sounds are heard:
- try to locate the source of the sound. Take the lift unit out of service and contact your Handicare dealer.
Symbol indication
The symbol indicates that the lift unit must be inspected. Contact your dealer.
The symbol lights up after 6-12 months.
Indicates battery charge status.
Emergency lowering battery
- Light up when accessory with 9V battery are connected and the power in the 9v battery is low
(need to be changed). Both symbols will light amber. Sound should indicate (pip) every minute.
In-rail charging
- When book symbol light and the battery symbol flashes the charging function is not functioning.
No contact with the charger.
Overload protection
- Prevents the lift from lifting more than the maximum load. If this happens A symbol for overload
and a warning triangle light up (amber) a signal is heard. If this symbol is activated, the user is too
heavy.
18
M A N U A L
System
RoMedic