Operation
28
S4600-ST®
6.11
"Pitot Measurement" program (option)
Starting the "Pitot Measurement" program
When you start the "Pitot Measurement" program, the pressure
sensor is first zero-balanced; this takes a few seconds.
The colour of the pressure value then changes from grey to black.
The black colour indicates that the device is ready for
measurements. Zero balancing can also be performed manually via
the main menu.
The main display shows the following measured values with
adjustable units:
Flow (m/s, km/h)
Volume (m³/h, l/s, m³/s)
Pressure (hPa)
Barometric pressure (hPa)
The optional Data Logger function is also available; it allows you to
save a series of measured values to the MicroSD card. The Data
Logger function is described in chapter 6.7, page 22.
Key
Function
Quitting the program "Pitot Measurement".
Changing selection in the menu.
Displaying submenu.
Switch off the device.
Entering measurement data (units, K factor of the Pitot tube,
chimney shape, chimney size)