Total no. of pages: 30
PAGE 28
Anleitung EAG80_26_00_2017_03_28_GB
Version 28.04.2017
8.4
Settings for scales settling time
Settings under 4.3 Menu filter settings (
FILtE
FIL
Activating scales settling time
AuS
Scales settling time deactivated (SF, FIL A and SPrun will be ignored)
EIn
Scales settling time activated (settings via SF, FIL A and SPrun)
8.5
Test scales with highest possible weight
Settings under 4.4.6 JUSt
If the weight value shown does not correspond to the weight on the scales, the scales should be adjusted
with weights.
Press and hold 0/ESC and Start/Stop Stop simultaneously until ALLGE appears in the display.
WAAGE
JUSt
Scales adjustment with weights
0-JuS
Empty platform and press Start/Stop
G-JuS
Place weight and press Start/Stop
00000
Enter weight value and press Start/Stop
JUSt
Adjustment completed