DSC Q Series Getting Started Guide
52
•
Autosampler Calibration via Instrument Control
a.
Select
Calibrate/
Autosampler
from the
menu. The window
shown in the figure to
the right will be dis-
played.
b. Click the plus (+) sign to
display the types of cal-
ibration available.
2.
Select the desired type of calibration from the list available.
•
To calibrate both the sample tray and cell areas select
Full auto cal
.
•
To calibrate the sample tray only select
Tray auto cal
.
•
To calibrate the cell areas only select
Cell auto cal
.
•
If you need to perform more refined calibrations, select any of the manual calibration functions.
See the online help for detailed information.
3.
Make sure you press the
Save
button when you have finished the desired calibration function.
After calibration is complete, you can perform a
Cycle Test
to test the calibrated positions.