DSC Q Series Getting Started Guide
53
Q1000/Q100 Autosampler Calibration
To perform Autosampler calibration from the touch screen follow the instructions below.
1.
Touch the
AUTOSAMPLER
button on the
CON-
TROL MENU
touch screen. The screen shown in the
figure to the right will be displayed.
2.
Select the desired type of calibration from the buttons
available.
•
To calibrate both the sample tray and cell areas
select
AUTO CAL
.
•
To calibrate the sample tray only select
TRAY
CAL
.
•
To calibrate the cell areas only select
CELL CAL
.
•
If you need to perform more refined calibrations,
select
MANUAL CAL
to access more calibration
functions.
See the online help for detailed information.
3.
Make sure you save the calibrations when you have finished the desired calibration function.
After calibration is complete, you can perform a
Cycle Test
to test the calibrated positions.