5
Performance Checks
Test Procedures
5-33
Table
Chapter 5-18, Frequency Measurement Accuracy
Function Generator
Setting
Error Limits
5251
Counter Reading
Pass
Fail
1.000000 MHz
±
2 Hz
100.0000 MHz
±
100 Hz
120.0000 MHz
±
200 Hz
2. Change the display time to Hold
3. Press the Reset/Arm button and verify that the frequency
reading is 120.000000 MHz,
±
200 Hz
Test Results
Pass
Fail
Period, Period
Averaged
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency:
As required by the test
Wave:
Square
Amplitude 500
mV
2. Connect the function generator to the 5251 TRIG IN connector
3. Configure the 5251, as follows:
Auxiliary Function: Counter/Timer
Function: Period
Trigger Level:
0 V
Test Procedure:
1. Perform Period Accuracy tests using Table 19
Table
Chapter 5-19, Period Measurement Accuracy
Function Generator
Setting
Error Limits
5251
Counter Reading
Pass
Fail
10 kHz
100.0
μ
s
±
100 ns
100 kHz
10.00
μ
s
±
100 ns
1 MHz
1.000
μ
s
±
100 ns
2. Change the counter/timer function to Period Averaged
3. With the last function generator setting in Table 27, verify that
the period reading is 1.000000
μ
s
±
50 ps
Test Results
Pass
Fail
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...