TE5251
User Manual
6-4
Recommended
Test Equipment
Recommended equipment for adjustments is listed in Table 6-1.
Instruments other than those listed may be used only if their
specifications equal or exceed the required minimal characteristics.
Also listed below are accessories required for calibration.
Table
6-1, Recommended calibration for Adjustments
Equipment
Model No.
Manufacturer
Oscilloscope (with jitter package)
LC684
LeCroy
Digital Multimeter
2000
Keithley
Frequency Counter (Rubidium reference) 6020R
Tabor Electronics
Function Generator (with manual trigger)
8571
Tabor Electronics
Accessories BNC
to BNC cables
50
Ω
Feedthrough termination
Dual banana to BNC adapter
Adjustment
Procedures
Use the following procedures to calibrate the Model 5251. The
following paragraphs show how to set up the instrument for calibration
and what the acceptable calibration limits are.
Calibration requires that ArbConnection utility be installed and
interfaced to the instrument.
Calibration is performed from the Calibration Panel in ArbConnection.
To invoke this panel, one requires a password that is available to
service centers only. Contact your nearest Tabor service center for
information and permit to obtain your calibration password. Use the
following procedure to calibrate the generator:
1. Invoke
ArbConnection
2. Click on the Calibration tab on the Panels bar
3. Expect to be prompted with the following dialog box
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...