6
Adjustments and Firmware Update
Pulse Response Adjustments
6-19
2. Connect the 5251 output to the DMM input. Terminate the
5251output at the DMM input with the, 50
Ω
Feed through
termination
3. Configure the 5251as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
1
V
Mode:
Modulation
Adjustment:
1. Adjust CAL:SETUP26 for DMM reading of 353.5 mV
±
3 mV
Pulse Response
Adjustments
The pulse response adjustments assure that the rise and fall times, as
well as, the aberrations are within the specified range. Use this
procedure if you suspect that the pulse response is an issue. Note that
setups 27 and 28 are marked as a factory adjustment and therefore, it
is not normally required to be performed during normal calibration
cycles except if the pulse response of the output stage has been
degraded and does not meet the published specification limits, or after
a repair has been executed on these very circuit.
(Setup 27)
Pulse Response, Amplifier Out
Equipment: Oscilloscope, BNC to BNC cable, 20 dB Feedthrough
attenuator
Preparation:
1. Configure the 5251as follows:
Function:
Square
Amplitude:
1
V
2. Connect the 5251 output to the oscilloscope input. Set
oscilloscope input impedance to 50
Ω
3. Set oscilloscope vertical sensitivity to 20 mV
Adjustment:
1. Adjust vertical trace to 6 divisions
2. Adjust RV1 for best pulse response (4 ns type, 5 % aberrations)
(Setup 28)
Pulse Response, Amplifier In
Equipment: Oscilloscope, BNC to BNC cable, 20 dB Feedthrough
attenuator
Preparation:
1. Configure the 5251as follows:
Function:
Square
Amplitude:
6
V
2. Connect the 5251 output to the oscilloscope input. Use 20 dB
Feedthrough attenuator at the oscilloscope input
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...