6
Adjustments and Firmware Update
Flatness Adjustments
6-27
Setup 47
Frequency Flatness – Modulation
Equipment: Oscilloscope, BNC to BNC cable, 20 dB Feedthrough
attenuator
Preparation:
1. Configure the 5251as follows:
Function:
Modulation
ON
Modulation:
Sweep
Start Freq:
1 MHz
Stop Freq:
100 MHz
Sweep Time:
1 ms
Marker:
1
MHz
Amplitude:
6
V
2. Connect the 5251 output to the oscilloscope input. Use 20 dB
Feedthrough attenuator at the oscilloscope input
3. Set oscilloscope input impedance to 50
Ω
4. Set oscilloscope vertical sensitivity to 0.1 V
Adjustment:
1. Adjust C10 for the best flatness.