9100A/9200A/9400
Maintenance and Performance Checks 3-10
Amplitude: 1
V/div
2. Connect the 9400 Channel 1 output monitor to the Oscilloscope
input.
3. Configure the Waveform Generator as follows:
Amplitude: 8
V
Function:
Square wave
Output: On
Frequency: 50
kHz
4. Connect the waveform generator to Channel 1 input. Terminate
the waveform generator using the 50
Ω
feedthrough terminator at
the 9400 input
Test Procedure
1. Using the variable vertical adjustment on the oscilloscope,
adjust the vertical trace to show exactly 6 vertical divisions
2. Perform rise/fall time tests on channel 1 output using Table 3-
10
3. Use the same procedure to check channels 2, 3 and 4
Table
3-10, Output Monitor Aberrations Tests
Parameter
Oscilloscope
Reading
Tested Error
Limits
CH1
CH2
CH3
CH4
Pass
Fail
Rise Time
<1
μ
s
Fall Time
<1
μ
s
Distortion Tests
Equipment
:
Distortion Analyzer, Arbitrary Waveform Generator, 50
Ω
feedthrough terminator
Preparation:
1. Connect the 9400 Channel 1 output monitor to the distortion
analyzer input.
2. Configure the Waveform Generator as follows:
Amplitude: 8
V
Function:
Sine wave
Output: On
Frequency:
As required for the tests
3. Connect the waveform generator to Channel 1 input. Terminate
the waveform generator using the 50
Ω
feedthrough terminator at
the 9400 input
Test Procedure
1. Perform Aberrations tests on channel 1 output using Table 3-11
2. Use the same procedure to check channels 2, 3 and 4