9100A/9200A/9400
Maintenance and Performance Checks 3-12
Gain Accuracy Tests
Equipment
:
DMM, Arbitrary Waveform Generator, Load Resistance,
50
Ω
feedthrough terminator
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 200V
2. Connect the 9400 Channel 1 output to the DMM input. Attach
the load resistance at the input terminals of the DMM
3. Configure the Waveform Generator as follows:
Frequency: 1kHz
Output: On
Amplitude:
As required for the test
4. Connect the waveform generator to Channel 1 input. Terminate
the waveform generator using the 50
Ω
feedthrough terminator at
the 9400 input
Test Procedure
1. Perform Gain Accuracy tests on channel 1 and channel 2
outputs using Table 3-12
2. Use the same procedure to check channels 3 and 4
Table
3-12, Unipolar Mode Gain Accuracy Tests
Arb Amplitude
DMM Reading
Setting Error
Limits
CH1
CH2
CH3
CH4
Pass
Fail
2 Vp-p
19.30 V,
±
2 V
4 Vp-p
38.45 V,
±
4 V
6 Vp-p
57.95 V,
±
6 V
8 Vp-p
76.80 V,
±
8 V
Bandwidth
Equipment
:
Oscilloscope, Arbitrary Waveform Generator, Load
Resistance, x100 high voltage probe, 50
Ω
feedthrough terminator
Preparation:
1. Configure the Oscilloscope as follows:
Time Base:
200
μ
s/div
Amplitude: 20
V/div
2. Connect the 9400 Channel 1 output to the load resistance
3. Connect the x100 high voltage probe across the load resistance
4. Configure the Waveform Generator as follows:
Amplitude:
8 V
Output: On
Frequency:
As required for the tests
Test Procedure
1. Using the variable vertical adjustment on the oscilloscope,
adjust the vertical trace to show exactly 6 vertical divisions
2. Perform bandwidth, large signals tests on channel 1 and
channel 2 outputs using Table 3-13
3. Use the same procedure to check channels 3 and 4