User Manual 9250
Portrayal 3-11
CH1 High Frequency
Response
Equipment:
Oscilloscope, waveform generator, x10 attenuator
Procedure:
50. Configure the waveform generator as follows:
51.
Function:
Square wave
52.
Frequency:
1MHz
53.
Amplitude:
1.6Vp-p (may vary depending on the
configured gain)
54. Connect the waveform generator output to the 9250 channel
1 input
55. Connect the 9250 channel 1 through the x10 attenuator to the
oscilloscope
56. Set oscilloscope to dc coupling and 50
Ω
termination
Adjustment:
1. Adjust RV5, RV8 and C77 for transition time between 18-20ns
and lowest aberrations
CH2 Low Frequency
Response
Equipment:
Oscilloscope, waveform generator, x10 attenuator
Procedure:
57. Configure the waveform generator as follows:
58.
Function:
Square wave
59.
Frequency:
1MHz
60.
Amplitude:
1.6Vp-p (may vary depending on the
configured
gain)
61. Connect the waveform generator output to the 9250 channel
2 input
62. Connect the 9250 channel 2 through the x10 attenuator to the
oscilloscope
63. Set oscilloscope to dc coupling and 50
Ω
termination
Adjustment:
1. Adjust RV1, RV2 and C35 for transition time between 18-20ns
and lowest aberrations